Precision Medical 7MFA1001 Quick Start Manual Download Page 2

SPECIFICATIONS

Model

Gas

Max Flush 

Flow

Increments

(lpm)

7MFA1001

Oxygen

40-60 lpm

.5, 1, 2, 3, 4, 5, 6, 8, 10, 

15 Flush

7MFA2001

Air

7MFA1001TC

Oxygen

40-60 lpm

.5, 1, 2, 3, 4, 5, 6, 8, 10, 

15 Flush

7MFA2001TC

Air

7MFA1101

Oxygen

N/A

1, 2, 3, 4, 6, 8, 10, 12, 15, 

20, 25

7MFA2101

Air

7MFA1101TC

Oxygen

N/A

1, 2, 3, 4, 6, 8, 10, 12, 15, 

20, 25

7MFA2101TC

Air

Flowmeters calibrated at 50 psi (3.4 bar), 70°F (21°C), standard 

atmospheric pressure.

Accuracy:

   ±10% of the indicated flow.

Operating Temperature:

  69°F (20.6°C) to 71°F (21.6°C)

Storage Temperature:

   -40°F (-40°C) to 140°F (60°C)

Max Humidity:

   95%

NOTE:

 Storage / Transport outside the specified range may cause 

damage to the flowmeter.

Specifications are subject to change without notice.

Tell us how we are doing!

Visit us at www.precisionmedical.com

OPERATING INSTRUCTIONS

NOTE: 

Precision Medical, Inc.

 

recommends the use of kink proof 

patient tubing.

1. Be sure Dial Flowmeter is in the “OFF” position.
2. Connect the Dial Flowmeter to a 50 psi (3.4 bar) Oxygen or Air 

Source.

3. Set flow by aligning the Indicating Pointer   to the desired flow.

NOTE:

 The accuracy of flow will be adversely affected by any 

variation of inlet pressure, ambient temperature or outlet 

resistance from rated values.

WARNING

•  Read this User Manual before installing or operating the Dial 

Flowmeter.

•  There is NO FLOW between settings. The Indicating Pointer   

must point to a specific number on the dial to obtain flow.

CAUTION

ONLY use appropriate indexed fittings to connect the Dial 

Flowmeter to appropriate source. Use 

Oxygen

 connections for 

Oxygen

 Flowmeters and use 

Air

 connections for 

Air

 Flowmeters. 

CLEANING / MAINTENANCE

1. Disconnect all connections before cleaning.
2. Clean exterior surfaces of the Dial Flowmeter with a cloth 

dampened with water.

3. Wipe dry with a clean cloth.
4. Store the Flowmeter in a clean area free from grease, oil and other 

sources of contamination.

CAUTION

•  DO NOT

 immerse the Dial Flowmeter in any kind of liquid. 

This will cause damage to the Dial Flowmeter and will void the 

warranty.

•  This Dial Flowmeter is not field serviceable.

CAUTION

Missing or illegible labels must be replaced, contact 

Precision Medical, Inc.

COMPONENT IDENTIFICATION

Indicating 

Pointer

Inlet 

Connection

Flow 

Increments

Outlet

Connection

7MFA1001

 

(shown)

TROUBLESHOOTING

If the Dial Flowmeter fails to function, consult the Troubleshooting Guide 

below. If the problem cannot be solved, consult your Provider.

Problem

Probable Cause

Remedy

No Flow

• Flowmeter set in 

“OFF” position

• Dial set between 

settings

• Internal failure

• Turn to desired 

setting

• Move Dial to flow 

setting

• Return to Precision 

Medical, Inc.

Dial will not turn / 

Hard to turn

• Dial seized

• Return to Precision 

Medical, Inc.

Will not shut off

• Flowmeter not in 

“OFF” position

• Turn Dial to the “OFF” 

position

RETURNS

Returned products require a Returned Goods Authorization (RGA) 

number. Any product returned to Precision Medical, Inc. must be 

packaged in a sealed container to prevent damage. Precision Medical, 

Inc. will not be responsible for goods damaged in transit. Return Policy 

available at www.precisionmedical.com.

Manuals are available on our website: www.precisionmedical.com.