WARNING

•  Use this Flowmeter only for its “Intended Use” as described 

in this manual.

•  alWaYS confirm prescribed flow before administering to 

patient and monitor flow on a frequent basis.

•  This Flowmeter contains magnetic, ferrous material that may 

affect the results of an MRI.

To Reduce the Risk of Fire or Explosion:

  alWaYS follow aNSI and CGa standards for Medical Gas 

Products and Flowmeters (E-7) and Oxygen Handling (G-4).

  DO NOT use oils, greases, organic lubricants or any 

combustible materials on or near this Flowmeter.

• 

DO NOT use near any type of flame or flammable/explosive 

substances, vapors or atmosphere.

• 

DO NOT smoke in an area where oxygen is being administered.

CAUTION

•  This Flowmeter must be operated with the Flow Tube in a 

vertical, upright position.

•  Only personnel instructed and trained in its use should 

operate this Flowmeter.

•  Be sure all connections are tight and leak free.

•  Only use oxygen-safe leak detector.

•  

DO NOT autoclave.

•  

DO NOT gas sterilize with EtO (Ethylene Oxide).

• 

DO NOT clean with aromatic hydrocarbons.

• 

DO NOT immerse product in any kind of liquid. This will void 

the warranty.

•  This Flowmeter may have a factory installed restrictor. Prior to 

use, check product labeling for flow restrictions.

•  This Flowmeter contains a glass flowtube which is fragile. 

Special care should be observed to avoid braking the flowtube.

Summary of Contents for 1MFA3001

Page 1: ...tative EMERGO EUROPE INC Molenstraat 15 2513 BH The Hague Netherlands SAVE THESE INSTRUCTIONS CAUTION Federal USA law restricts this device to sale by or on the order of a physician 300 Held Drive Tel...

Page 2: ...Flowmeter This is provided for your safety and to prevent damage to the Flowmeter If you do not understand this manual DO NOT USE the Flowmeter and contact your Provider safety information warnings an...

Page 3: ...xplosive substances vapors or atmosphere DO NOT smoke in an area where oxygen is being administered CAUTION This Flowmeter must be operated with the Flow Tube in a vertical upright position Only perso...

Page 4: ...lpm 20 40 lpm Transport Storage Requirements 40 F 40 C to 140 60 C Restricted Max Flush Flow Note Storage Transport outside the specified range may cause damage to the flowmeter The effect on accuracy...

Page 5: ...ow Tube NOTE If the Float is not resting at the bottom of the Flow Tube the product is leaking consult the Troubleshooting Guide 4 Adjust Flow To increase Turn Knob counterclockwise To decrease Turn K...

Page 6: ...and water 3 Wipe dry with a clean cloth Troubleshooting If the Flowmeter fails to function consult the Troubleshooting Guide below If problem cannot be corrected consult your Provider or Precision Med...

Page 7: ...n Representative Emergo Europe Molenstraat 15 2513 BH The Hague The Netherlands Product Flowmeters Model s 1MFA MDD Class IIb Classification criteria Clause 3 2 Rule 11 of Annex IX of MDD As delivered...

Page 8: ...l Inc or any retailers are not authorized to make oral warranties about the merchandise described in this contract and any such statements shall not be relied upon and are not part of the contract for...