ProtEX-MAX PD8-6310-WM Explosion-Proof NTEP Certified Batch Controller Instruction Manual
11
Compliance Information
Safety
UL & c-UL Listed
USA & Canada
UL 508 Industrial Control Equipment
UL File Number
E160849
Front Panel
UL Type 4X, NEMA 4X, IP65; panel gasket provided
Low Voltage
Directive
EN 61010-1:2010
Safety requirements for measurement, control, and laboratory use
NTEP Weights and Measures Certification
Certificate Number
14-061
Definition
Register for Meter
Stationary Wholesale Liquid Register/Controller
Digital Electronic
Model: PD63XX Series and PD8-63XX Series
Evaluation Criteria
NIST Handbook 44 Specifications, Tolerances, and Other Technical
Requirements for Weighing and Measuring Devices, 2014 Edition. NCWM
Publication 14 Measuring Devices, 2014 Edition.
National Conference on Weights and Measures (NCWM) National Type Evaluation Program (NTEP)
Certificate of Conformance available at www.predig.com
Electromagnetic Compatibility
Emissions
EN 55022:2010
Class A ITE emissions requirements
Radiated
Emissions
Class A
AC Mains
Conducted
Emissions
Class A
Immunity
EN 61326-1:2013
Measurement, control, and laboratory equipment
EN 61000-6-2:2005
EMC heavy industrial generic immunity standard
RFI - Amplitude
Modulated
80 -1000 MHz 10 V/m 80% AM (1 kHz)
1.4 - 2.0 GHz 3 V/m 80% AM (1 kHz)
2.0 - 2.7 GHz 1 V/m 80% AM (1 kHz)
Electrical Fast
Transients
±2kV AC mains, ±1kV other
Electrostatic
Discharge
±4kV contact, ±8kV air
RFI - Conducted
10V, 0.15-80 MHz, 1kHz 80% AM
AC Surge
±2kV Common, ±1kV Differential
Surge
1KV (CM)
Power-Frequency
Magnetic Field
30 A/m 70%V for 0.5 period
Voltage Dips
40%V for 5 & 50 periods
70%V for 25 periods
Voltage
Interruptions
<5%V for 250 periods