Model PD750, PD751, PD752 Universal Temperature Meters
Instruction Manual
RECOMMENDED CALIBRATION POINTS
SCALE
(
F or [
)
F or C may be switched on to indicate Fahrenheit or Celsius.
Selecting Fahrenheit or Celsius
1. Press
ENTER
, when
F or [
appears press
ENTER
again.
2. The meter will display
F or [
alternately. Press
ENTER
when desired character is
flashing. Meter will then return to indication mode and display the temperature in
either Fahrenheit or Celsius.
12
www.predig.com
Type
Type of Input
Input 1
Input 2
Check
Number
(Low)
(High)
(Middle)
0
Type J T/C
32°F
1182°F
600°F
1
Type K T/C
32°F
1183°F
960°F
2
Type T T/C
32°F
693°F
360°F
3
Type T T/C
32.0°F
693.0°F
360.0°F
4
Type E T/C
32°F
1652°F
840°F
5
Type R T/C
100°F
3000°F
1500°F
6
Type S T/C
100°F
3000°F
1500°F
7
100Ω RTD
32.0°F
1148.0°F
590.0°F
(0.00385, 0.1°)
100Ω
320.12Ω
215.61Ω
8
100Ω RTD
32.0°F
1127.0°F
580.0°F
(0.00392, 0.1°)
100Ω
320.89Ω
215.87Ω
9
100Ω RTD
32°F
1148°F
590°F
(0.00385, 1°)
100Ω
320.12Ω
215.61Ω
10
100Ω RTD
32°F
1127°F
580°F
(0.00392, 1°)
100Ω
320.89Ω
215.87Ω