™ PD6100 Strain Gauge, Load Cell, and mV Meter
Instruction Manual
19
Setup and Programming
There is
no need to recalibrate
the meter
when first received from the factory.
The meter is
factory calibrated
prior to
shipment for millivolts with calibration
equipment that is certified to NIST standards.
Overview
There is a 5-position DIP switch to set the meter input
selection. See
setup the switch.
Setup and programming is done through the front
panel buttons.
After power and input signal connections have been
completed and verified, apply power to the meter.
Front Panel Buttons and Status
LED Indicators
Button
Symbol
Description
LED
Status
Menu
1-8
Alarm 1-8
indicator
Right
arrow/F1
Zero meter
1-8
M
Flashing:
Relay in
manual
control mode
Up arrow/F2
Reset tare
T
Flashing:
Tare
Enter/F3
Tare
G
Gross
Note:
F4 is a digital input. Alarms
5-8 are enabled when relay
expansion module is
installed.
1-8
Flashing:
Relay
interlock
switch open
Note:
LEDs for relays in
manual mode flash
with the “M” LED every
10 seconds. “M”
flashing by itself
indicates Aout
–
manual control is used.
•
Press the Menu button to enter or exit the
Programming Mode at any time.
•
Press the Right arrow button to move to the next
digit during digit or decimal point programming.
•
Press or hold the Up arrow button to scroll
through the menus, decimal point, or to
increment the value of a digit.
•
Press the Enter button to access a menu or to
accept a setting.
•
Press and hold the Menu button for three
seconds to access the advanced features of the
meter.