Helios PD2-6363 Dual Pulse Input Rate/Totalizer
Instruction Manual
34
Scaling the Meter for Channel A
(
SCal-A
)
For instructions on how to program numeric values
Note: The display values (
DiS 1
and
DiS 2
) need to
be in units of measure per second. For example, if the
rate at 1000 Hz is 13.20 gal/min, this must be
converted to gal/sec. In this scenario, the numbers
input into the
SCALE
menu would be as follows:
InP 1
: 0.0
DiS 1
: 0.00
InP 2
: 1000.0
DiS 2
: 0.22
Set the time base set to minutes (see
) on page
) and the meter will
display 13.20 gal/min at 1000 Hz
Error Message (
Error
)
An error message indicates that the calibration or
scaling process was not successful.
After the error message is displayed, the meter
reverts to input 2 during calibration or scaling and to
input 1 during internal calibration, allowing the
appropriate input signal to be applied or programmed.
The error message might be caused by any of the
following conditions:
1. Input signal is not connected to the proper
terminals or it is connected backwards.
2. Minimum input span requirements not
maintained.
3. Input 1 signal inadvertently applied to calibrate
input 2.
Minimum Input Span
The minimum allowed input span is 1.0 Hz, which is
the minimum difference between input 1 and input 2
signals required to complete the calibration using an
external signal source, or 0.1 Hz for scaling the meter
without a signal source.
Calibrating the Meter with External
Source (
CAL-A
,
CAL-B
)
To scale the meter without a signal source, refer to
Scaling the Meter without a Signal Source
), page
The meter can be calibrated to display the process
variables in engineering units by applying the
appropriate input signal and following the calibration
procedure.
The use of a calibrated signal source is strongly
recommended to calibrate the meter.
There is no need to warm up the meter before
performing calibration of the inputs.
Channel B is calibrated similarly to Channel A, shown
below.
Calibrating the Meter for Channel
A (
CAL-A
)
Note: The display values (
DiS 1
and
DiS 2
) need to
be in units of measure per second.
Note: Inputs for the above example are:
Input 1: 0.0 Hz; Display 1: 0.0 gallons/min
Input 2: 1000.0 Hz; Display 2: 6000.0 gallons/min