MANUFACTURER

Type of Information

Specifications

Device Name

Powered Impression Gun

Model Number

BFC

3

Battery

11.1V lithium-ion

Battery charger

12.6V/500mA with LED indicator

Charging time

2

1

2

 hours (full charge)

Flow rate

Adjustable from 0.6 to 1.7ml/sec.

Device Dimensions

4"W x 4"D x 7"H (overall)

Device Weight

28.1 oz. 

Protection from  
Ingress of Liquid

IPX0

PRODUCT SPECIFICATIONS

CHARGING THE BATTERY

Use only the charger that came with your BFC

3

. Use of 

another charger could damage the unit and will void 
the warranty.

With a fully charged battery, the BFC

3

 can dispense 

about 60 full 50ml heavy-body VPS cartridges. When 
battery output is reduced, the low battery LED 

K

 

will illuminate, indicating that the BFC

3

 battery should 

be charged soon. The BFC

3

 will dispense up to 10 full 

cartridges after the low battery LED illuminates. 
The BFC

3

 battery is rated for 500 recharges (about 

30,000 cartridges) before replacement will be required. 
Contact Practicon for battery replacement service. Do 
not attempt to open the unit or service the battery.

CLEANING AND DISINFECTION

This product must be disinfected for each patient. 
Failure to follow these guidelines could damage the 
device and void the warranty. 

Clean the body of the device and plungers with disin-
fectant wipes. Follow all disinfectant manufacturer’s 
guidelines for use.
•  Do not use any cleaner containing bleach.
•  Do not immerse in liquid.
•  Do not place in autoclave.

PREPARING THE BFC

3

 FOR USE

Select the 50ml plunger 

C

 or 75ml plunger 

D

 as 

appropriate for your material cartridge. Open the 
cartridge clamp 

G

. Hold the BFC

3

 in one hand and 

press upward on the plunger release lever 

M

 with your 

thumb. With the other hand, insert the plunger 

teeth 

side down

 into the plunger slot 

L

 in the front of the 

unit and slide it back as far as it will go. Release the 
plunger release lever. The BFC

3

 is now ready to use.

OPERATION

Using the BFC

3

 is very much like using a manual 

impression gun, only easier. Instead of having to 
repeatedly squeeze a large spring-loaded trigger with 
your hand, BFC

3

 dispenses automix material with the 

press of a button.

LOADING A CARTRIDGE

Press the plunger release lever 

M

 upward with your 

thumb, pull the plunger back until it stops, and then 
release the lever. Open the cartridge clamp 

G

 

and 

insert an automix cartridge. Close the cartridge clamp. 

Press the plunger release lever again and slide the 
plunger forward until it contacts the cartridge.  
Release the lever.

ADJUSTING THE FLOW RATE

The BFC

3

 can dispense material at a rate that ranges 

from 0.6 to 1.7ml/sec. Adjust the flow rate by rotating the 
dispensing speed control 

J

 between the MIN and MAX 

settings. Experiment with different flow rates to find the 
setting that works best for your practice and procedures.

DISPENSING MATERIAL

As with all impression guns, express the first drops of 
material onto a paper towel or tray liner.

Press the dispensing activation trigger 

H

 with your 

index finger to start the flow of material. Release the 
trigger when enough material has been dispensed. 
The BFC

3

 includes an anti-drip feature that momentarily 

reverses the plunger when the dispensing trigger 
is released, and an auto-stop feature with overload 
protection when a cartridge has been emptied.

Problem

Resolution

Cannot insert plunger

Insert the plunger into the slot on the front of the unit 

teeth side down

The slot is keyed so that the plunger can only go in one way.

Dispensing is too slow/fast

Adjust the dispensing speed control to increase/decrease the flow.

Unit will not dispense material

Cartridge is empty and auto-stop engaged; replace cartridge.  
Battery charger is still connected to unit (by design, unit will not oper-
ate with charger connected); disconnect battery charger from BFC

3

.

Battery will not charge

Contact Practicon for battery service. Do not attempt to open the unit 
or service the battery.

TROUBLESHOOTING

Product Information 

BFC

3

 COMPONENTS AND CONTROLS

J

 

Dispensing speed control 

Adjustable from 0.6ml/sec. (min)  

to 1.7ml/sec. (max)

L

 

Plunger slot

N

 

Battery charger connector

H

 

Dispensing activation trigger

G

 

Cartridge clamp 

Raise to open and insert/remove 

automix cartridge

M

 

Plunger release lever  

Press upward to insert, withdraw, 

or remove plunger

K

 

Low battery LED

G

L

J

K

N

H

M

CHARGE THE BATTERY FULLY BEFORE USING THE BFC

3

 FOR THE FIRST TIME

Insert the charging cable into the connector 

N

 at the base of the BFC

3

 handle and then place the BFC

3

 into the 

stand so the charging cable extends out the back of the stand. The LED indicator on the wall charger will change 
from red to green when the battery is fully charged. A full charge can take up to 2

1

2

 hours.

DISCONNECT CHARGER FROM BFC

3

 PRIOR TO USE.  

UNIT WILL NOT OPERATE WITH BATTERY CHARGER CONNECTED.