5
Alarm Set Point Adjustment
The dew point at which an alarm occurs can be changed. The unit must be opened for
this procedure so
remove power before proceeding.
Remove the four nuts and screws that hold the panel in the case. Remove the panel and
look at the PC board.
WARNING
THE FOLLOWING PROCEDURE IS PERFORMED WITH POWER APPLIED. TO
PREVENT SHOCK, DO NOT TOUCH ANY TERMINALS INSIDE THE DEW
POINT MONITOR.
Apply power to the unit and turn it on with care, making sure not to touch any electrical
components inside. Locate the display function switch (S1) near the upper left corner of
the PC board. Note that only one position of this switch should be ON at a time.
Switch the OPER position of S1 to OFF and the SET1 position to ON. Adjust SET1
potentiometer to the desired alarm set-point as shown on the LED display.
S1 setting for alarm set-point view and adjustment
When finished adjusting the alarm set-point, return S1 to the default operating mode by
switching OPER position ON and all others OFF as shown below:
S1 setting for normal operation
Remove power to the unit before assembling the panel back in its case.
OPER
SET1
SET2
S1
OPER
SET1
SET2
S1