TARGAS-1 Operation Manual V. 1.02
70
SRC – Start Process (Step 1)
The first screen asks the user to confirm that they would like to start the SRC process.
SRC – Start Process
Back Button
Returns to the Processes Menu.
Right Arrow
Button
Confirms and starts the SRC process by displaying the SRC – Volume and Area
Settings screen (Step 2).
SRC – Volume and Area Settings (Step 2)
The volume and area settings are defined in this screen.
SRC – V and A Settings
Def Button
Sets the volume and area values back to the defaults.
Volume Button
Sets the volume of the system. Entries between 1,000 and 2,000 ml are
permitted. Default is 1171 ml.
Area Button
Sets the surface area of the chamber. Entries between 50 and 100 cm
2
are
permitted. The default value is 78 cm
2
.
Volume/Area
Ratio
This value is automatically updated when there is a change to either the Volume
or Area values.
Stop Button
Stops the SRC process and returns to the Processes Menu.
Right Arrow
Button
Continues to SRC – Termination Settings screen (Step 3).