Other Status Codes/Messages
Message
What it Means / Action
Divide by Zero
Possible corruption of system data pointers or calculation error.
Download any stored data from internal memory and then clear
the internal memory. If problem persists contact PP Systems.
Stored DIFF BAL CO2 out of range
The maximum reference CO
2
concentration has changed and is
significantly higher than the maximum reference CO
2
recorded
when you performed a Stored Diff Bal Calibration. This is
common when there is a change in desiccants after initial Stored
Diff Bal Calibration. Perform a new Stored Diff Bal Calibration
with the new desiccants.
Stored Diff Bal H2O out of range
The maximum reference H
2
O concentration has changed and is
significantly higher than the maximum reference H
2
O recorded
when you performed a Stored Diff Bal Calibration. This is
common when there is a change in desiccants after initial Stored
Diff Bal Calibration. Perform a new Stored Diff Bal Calibration
with the new desiccants.