Claw Medical Vacuum Systems

maintenance Schedule

operating Hours

item

action needed

500

2500

3000

5000

10,000

15,000

20,000 remarks

Inlet air filter

Replace

l

s

l

Every 3,000 hours or less

Blower fan

Clean

l

l

l

l

Fan shield

Clean

l

l

l

l

Pump fans

Clean

l

l

l

l

Vacuum switch

Inspect

l

l

l

l

Gear oil

Replace

l

s

Every 3,000 hours or less

 

l

 Inspect

  s

 Replace

noTeS:

  1. Inspect and perform maintenance periodically according to maintenance schedule.
  2. The maintenance schedule relates to the normal operating conditions. If the circumstances and load condition are adverse, 

shorten the interval time and perform maintenance accordingly.

maintenance

lUBriCaTion

Refer to the manual from the vacuum pump manufacturer 
for lubrication intervals and type of oil to use. Oilless claw 
vacuum pumps have gear boxes that must be maintained and 
the gearbox oil needs to be changed at specified intervals.

 

If the gas being pumped  

  through the vacuum system 

is known to be incompatible with petroleum based oils, the 
user must drain and refill the vacuum pumps with suitable oil 
before use. Failure to do so will void the warranty.

Summary of Contents for CVPT Series

Page 1: ...not avoided WILL result in death or serious injury Warning indicates a potentially hazardous situation which if not avoided COULD result in death or serious injury Caution indicates a potentially haz...

Page 2: ...eiver Tank The receiver tank in a Powerex Medical Vacuum System is plumbed so that it sees the system vacuum It is not pressurized in normal use and should not be connected to any pressure sources The...

Page 3: ...ctrical supply before servicing the equipment Electrical shock hazard Make sure the system is grounded in accordance with NEC and local requirements All electrical hook ups must be performed by a qual...

Page 4: ...f the system extends beyond 600 equivalent straight feet use one size larger pipe Approximate system CFM equals the number of pumps running times CFM in table below Data below is for reference only if...

Page 5: ...he vacuum level is maintained and the minimum run timer conditions are satisfied 7 Check for excessive vibration unusual noises or leaks during operation If problems are detected shut down the system...

Page 6: ...maintenance schedule relates to the normal operating conditions If the circumstances and load condition are adverse shorten the interval time and perform maintenance accordingly Maintenance LUBRICATIO...

Page 7: ...eck drive coupling replace if needed 2 Verify that pump shaft turns freely repair or replace as necessary 3 Check vacuum discharge piping to verify air is being exhausted repair or replace as necessar...

Page 8: ...ns iii that have been materially modified without the prior written approval of Powerex or iv that experience failures resulting from operation either intentional or otherwise in excess of rated capac...

Reviews: