Product Manual: PXP-40HF Poskom Co.,
Ltd.
The kVp and mAs setting, memory selection and collimator lamp and laser pointer illumination
functions are controlled remotely by the hand switch by pressing the “Prep.” stage push-button as
follows:
<TRIPLE CLICK>
ATTENTION
Click the “Prep.” hand switch three (3) times consecutively. This will allow you
to enter in the “remote hand control” menu mode for operation.
11.3. MEMORY SELELCTION
When you click the PREP stage hand switch quickly three (3) times, the Pre-set Memory station LED
(PXP-20HF: 1-4 / PXP-40HF & 60HF: 1-8) will blink sequentially as well as the kVp and mAs LED
readouts. Make a single (1) click with the hand switch when the desired pre-set memory station lamp
begins to blink. This will cause the pre-set memory data to be selected and displayed in the control
LCD window.
NOTE
All pre-set memory data should be entered and stored at each memory station (PXP-20HF: 1-4 /
PXP-60HF: 1-8), prior to performing the above mode of operation.
11.4. kVp
SELELCTION
One (1) click of the “Prep” switch will cause the kVp LED readout to blink. You will now be in the kVp
selection menu mode. The kVp values increase by one (1) kVp unit per click of the “Prep.” switch. If
you depress and hold the prep switch for a few second, the kVp readout will increases rapidly up to
the maximum value (PXP-20HF: 90kVp / PXP-40HF & 60HF: 100kVp). The kVp readout will circulate
from lowest to highest values as long as you hold and press the button. When the desired kV readout
appears on the control display panel, release the prep hand switch and wait for a few seconds. The
kVp value will set up automatically.
13