POSITECTOR SHD Verifier Instruction Manual Download Page 3

1. Remove the PosiTector 

SHD Verifier

and Smart Sensor Indicator from the case.

Make sure to place the PosiTector 

SHD

Verifier

on a stable flat surface.

2. Connect  the  Load  Cell  Cable  to 

the Smart Sensor Indicator

The PosiTector

SHD Verifier

includes Certificates of Calibration for the SSI

and load cell. For information on returning your PosiTector

SHD Verifier 

for

recertification, contact DeFelsko

or your local dealer.

The PosiTector

SHD Verifier,

with valid Certificates of Calibration, can be

used  to  check  the  accuracy  of  PosiTector

SHD  Shore  Hardness

Durometers

. A sample Certificate of Accuracy is included.

  The SSI device comes pre-configured for verification of PosiTector

SHD

Shore Hardness Durometers

. No additional configuration is required or

recommended.

If the SSI is mistakenly altered, configuration can be restored by following

the steps outlined in our 

Reconfiguration Guide

. Contact DeFelsko for more

information.

2

3

PosiTector SHD Verifier Set-up

PosiTector SHD Verifier Set-up

Reconfiguring the SSI Device

Reconfiguring the SSI Device

Certification

Certification

3. Select a 

SHD A

(aluminum) or 

SHD D

(brass) target and insert it into

the load cell as shown. 

Note:

Inspect the targets before installing. To

prevent damage to the PosiTector 

SHD D

probe tip, the 

SHD D

(brass)

target is designed to be dented during testing and will periodically need to

be replaced. The 

SHD A

(aluminum) target will not be dented during testing

and should be replaced if there is any damage to the top surface.

4.1 Perform a Gage 

Reset

on the PosiTector 

SHD

4.2 Select 

Continuous  Reading  Mode

from  the  Setup  menu  on  the

PosiTector 

SHD

.

4.3 Turn the Micrometer Adjustment so that it is set between 1 and 2.

4.4 Place the PosiTector 

SHD

probe into Probe Holder. Ensure that the

probe tip doesn’t touch the target. 

For detailed instructions see the 

PosiTector 

SHD

Calibration Procedure

at dl.defelsko.com/pdf/manuals/PosiTectorSHD-Verifier-v1.0.pdf

Shore A

10

20

30

40

50

60

70

80

90

SHD Verifier 
Displays (lbs)

0.292 0.461 0.629 0.798 0.967 1.135 1.304 1.472 1.641

Shore D

10

20

30

40

50

60

70

80

90

SHD Verifier 
Displays (lbs)

0.999 1.999 2.998 3.997 4.996 5.996 6.995 7.994 8.994

4.5 Power on the Smart Sensor Indicator

4.6 Turn the Micrometer Adjustment to lower the probe onto the target until 

the desired load is displayed on the 

SHD Verifier

corresponding to the 

Shore A

or 

Shore D

verification points illustrated below.

4.7 Verify that the PosiTector 

SHD

is reading  1 Shore value of 

the corresponding load. 
Repeat steps 4.6 – 4.7 for each verification point.

Quick Calibration Procedure

Quick Calibration Procedure

SHD

Target

Probe

Holder

Target

Micrometer

Adjustment

www.

.com

[email protected]

1.800.561.8187

Reviews: