IMPORTANT:
READ MANUAL COMPLETELY BE F O R E O P E R A T I N G T H I S D E V I C E
Basic delivery technique is described. Also, this manual contains instructions on periodically required
checks to be performed by the user. These checks are necessary to insure the proper performance of
this device and its safety features. Retain this manual for future reference.
W A R N I N G S A N D P R E C A U T I O N S
These warnings and precautions are to help
you to understand how to safely operate the
Nitronox device. A WARNING alerts you to a
possible hazard to people. A CAUTION alerts
you to the possibility of equipment damage.
WARNING:
New or modified installations -
properly connected gas pipelines are
absolutely essential to patient safety.
The
ultimate responsibility of assuring that lines
are not crossed rests with the user. See
next page for details.
WARNING:
Porter equipment utilizes the
cross+protection
system. The flexible hose
and connectors that connect to the housing are
diameter indexed; 3/8” O.D. for Nitrous Oxide
and ½” O.D. for Oxygen. The
cross+protection
system is designed to
prevent misconnection of Oxygen and Nitrous
Oxide hoses.
DO NOT ATTEMPT TO
CHANGE THE DIAMETERS OR
CONNECTORS OF THE DEVICE!
Tampering
with the
cross+protection
system constitutes
acceptance of liability by the installer.
WARNING:
Nitronox® Inhalation Analgesia
systems are intended to be used by medical
personnel trained in its use and the use of
nitrous oxide and oxygen for medical
applications.
WARNING:
Do not use this device for the
administration of general anesthesia or as a
part of, or in conjunction with, a general
anesthesia administration system.
WARNING: Use Scavenging
Monitor for N2O in the patient treatment area
to insure that controls are effective in achieving
low levels of ppm (parts per million) exposure.
Contact Porter or Authorized Distributor for
details on monitors and testing.
Inspect and maintain the analgesia delivery
system to prevent N2O leaks in all hoses,
connections and fittings. Have all leaks
repaired immediately.
WARNING:
Medical workers are exposed to
N
2
O during administration of N
2
O/O
2
conscious
sedation analgesia. NIOSH has recommended
that exposures should be minimized. Contact
NIOSH (1-800-35-NIOSH) to receive NIOSH
Publications on
Control of Nitrous Oxide in
Dental Operatories.
Exposure can be
minimized by effective controls. National
Institute for Occupational Safety and Health
(NIOSH) publications state that controls,
including System Maintenance, Ventilation and
Work Practices can effectively reduce N
2
O
concentrations in patient procedures. Your
accessory Porter Nitronox Scavenger System
is an important part of the system of controls in
medical settings.
W A R N I N G :
W h e n u s i n g a s i ng l e u s e
f u l l f ac e m a s k , n o t s u p p l i ed s t e ri l e.
D i s p os e of m as k a f t e r u s e t o p r e v e n t
p a t i e nt c r os s - c on t a m i n at i o n. W h e n
u s i n g m u l t i p l e us e m as k , f o l l o w
m a n u f ac t ur e r ’ s s t e r i l i z at i on
i n s t r u c t i o n s .
CAUTION:
Always use clean, dry medical
grade gases. Introduction of moisture or other
contaminants into this device may result in
defective operation.
CAUTION:
Do not attempt to repair, alter or
calibrate this device. Unauthorized repair,
alteration or misuse of this device is likely to
adversely affect the performance and will void
the warranty.
CAUTION:
Never oil or grease any part of this
system (minimize fire or explosion potential).
2