IMPORTANT: 

READ MANUAL COMPLETELY BE F O R E   O P E R A T I N G   T H I S   D E V I C E  

Basic delivery technique is described. Also, this manual contains instructions on periodically required checks 
to be performed by the user.  These checks are necessary to insure the proper performance of this device 
and its safety features.  Retain this manual for future reference. 

 

 

WARNINGS AND PRECAUTIONS 

 

These  warnings  and  precautions  are  to  help  you 
to  understand  how  to  safely  operate  the  MXR 
Flowmeter.  A WARNING alerts you to a possible 
hazard  to  people.    A  CAUTION  alerts  you  to  the 
possibility of equipment damage. 

WARNING:

    Do  not  use  this  device  for  the 

administration  of  general  anesthesia  or  as  a  part 
of,  or  in  conjunction  with,  a  general  anesthesia 
administration system. 

WARNING:

  Workers  are  exposed  to  N

2

O  during 

administration  of  N

2

O/O

conscious  sedation 

analgesia.    NIOSH  has  recommended  that 
exposures should be minimized.  Contact NIOSH 
(1-800-35-NIOSH) to receive NIOSH Publications 
on 

Control of Nitrous Oxide in Dental Operatories.  

Exposure can be minimized by effective controls.  
National  Institute  for  Occupational  Safety  and 
Health  (NIOSH)  publications  state  that  controls, 
including  System  Maintenance,  Ventilation  and 
Work  Practices  can  effectively  reduce  N

2

concentrations  in  patient  procedures.    Your 
flowmeter  accessory  Porter  scavenger  system  is 
an important part of the system of controls. 

 

WARNING:

  Porter Instrument equipment utilizes 

the 

cross+protection 

system. The flexible hose 

and connectors that connect to the flowmeter are 
diameter indexed; 3/8” O.D. for Nitrous Oxide and 
½” for Oxygen. The 

cross+protection 

system is 

designed to prevent misconnection of Oxygen 
and Nitrous Oxide piping.  

DO NOT ATTEMPT 

TO CHANGE THE DIAMETERS OR 
CONNECTORS OF THE FLOWMETER!  

Tampering with the 

cross+protection

 system 

constitutes acceptance of liability by the installer. 
For your own protection, as well as that of the 
Doctor and the patients, use 3/8” O.D. tubing for 
all Nitrous O

xide lines and ½” O.D. tubing for all 

Oxygen lines.  

To assure safe operation and conformation to 
local fire codes, all Porter Instrument flowmeter 
systems are designed to be used with sedation 
delivery systems mounted inside walls and they 
meet or exceed the guidelines established by the 
National Fire Protection Association for 

Nonflammable Medical Gas Systems, NFPA 99. 
Copies of NFPA 99 or portions thereof may be 
obtained by writing to:  National Fire Protection 
Association, Batterymarch Park, Quincy, MA 
02269-9904; or call 1-800-344-3555 

WARNING:

  New or modified installations - 

properly connected gas pipelines are 
absolutely essential to patient safety.

  The 

authorized distributor or contractor should provide 
written documentation that all gas pipelines are 
connected properly and that the system has been 
pressure tested prior to use.  While this is a good 
business practice, it is important that the user 
verify by their own test, independent of the 
authorized distributor or contractor, that all gas 
pipelines are connected correctly prior to using 
the system.  

The ultimate responsibility of 

assuring that lines are not crossed rests with 
the user. 

WARNING:  

During any power outage, remember 

to turn OFF the flowmeter and manually turn OFF 
the tank valves.   With centralized, electrically 
powered gas systems, if gas was flowing when 
the power went out and the flowmeter is left ON, 
gas will be flowing when the power is restored. 

CAUTION:

    This  device  is  for  use  by  or  on  the 

order  of  a  physician,  dentist  or  licensed 
healthcare professional. 
 

CAUTION:

  Always  use  clean,  dry  medical  grade 

gases.    Introduction  of  moisture  or  other 
contaminants  into  this  device  may  result  in 
defective operation. 

CAUTION: 

Do  not  attempt  to  repair,  alter  or 

calibrate  this  device.    Unauthorized  repair, 
alteration  or  misuse  of  this  device  is  likely  to 
adversely affect the performance and will void the 
warranty. 

CAUTION:

 Never oil or grease any part of this 

system (minimize fire or explosion potential).

 

Summary of Contents for Matrx MDM

Page 1: ...ucts Division Mailing Address Riverside Road Pottington Business Park Barnstaple EX 31 1NP England Phone 44 0 1271 313131 Fax 44 0 1271 373636 The Quality System for Porter Instrument is certified to ISO 13485 The scope of our registration is The design manufacture distribution and servicing of Nitrous Oxide Oxygen Sedation Flowmeters Gas Scavenging Systems Gas Distribution Systems and Office Comm...

Page 2: ...tection as well as that of the Doctor and the patients use 3 8 O D tubing for all Nitrous Oxide lines and O D tubing for all Oxygen lines To assure safe operation and conformation to local fire codes all Porter Instrument flowmeter systems are designed to be used with sedation delivery systems mounted inside walls and they meet or exceed the guidelines established by the National Fire Protection A...

Page 3: ...Slide one end of breathing tube over outside diameter of breathing tube connector 3 Slide connector Y fitting of large nasal inhaler into free end of breathing tube Or N2O O2 Scavenger Large Scavenger and 3 Liter Breathing Bag Figure 2 Assembly Procedure 1 Position breathing bag as shown and slide opening in breathing bag over outside diameter of bag mount Figure 1 2 Slide the 22mm hose cuff of th...

Page 4: ...ks at pressure connections on the unit Bubbles will appear at leaking locations when a soap water solution is used This procedure is recommended each time a cylinder is changed Refer to FM 916 Mobile E Tank Stand Installation and Instructions for connecting gas supply to the flowmeter and tank installation to mobile stand MDM SPECIFICATIONS Figure 3 Physical Size 5 wide x 9 high x 12 deep Weight 7...

Page 5: ...MDM SPECIFICATION DIAGRAMS STANDARD MDM FLOWMETER HEADS Figure 3 4 ...

Page 6: ...urn on manifold medical gas system as per manufacturer s instructions Adjust line pressures as instructed Pressure must be 50 55 PSIG 5 For portable systems utilizing E cylinder tanks refer to FM 916 Mobile E Tank Stand Installation and Instructions The Controls and Indicators Mixture Dial controls setting of oxygen concentration of gas mixture Adjustment of this control does not affect total flow...

Page 7: ...ide flowmeter float falls at same rate as oxygen flowmeter float 5 Oxygen Flush Test Check Monthly Depress and release oxygen flush button to assure that breathing bag can be inflated 6 Non Rebreathing Valve Test Check Monthly Blow into the breathing tube connector Exhaled air must not pass through breathing connector 7 Air Inlet Valve Test Check Monthly Temporarily plug breathing bag mount and in...

Page 8: ...tion MAINTENANCE AND SERVICE It is advisable on a two 2 year cycle to have the MXR Flowmeter factory checked and serviced Perform Functional Tests Monthly as described on page 4 Inspect and maintain the analgesia delivery system to prevent N2O leaks in all hoses connections and fittings Repair all leaks immediately All service and repair must be accomplished at Matrx by Parker Have your authorized...