background image

Court  Results

1.

Press the 

MODE 

button and the message 

Court Results

will appear on the readout 

(Photo P).  This mode allows the user to recall the calibration average, baskets 1 and 
2 average, and compute the percentage difference between the two baskets. 

2.

Press the 

EXECUTION 

button and the calibration average as 

measured will be displayed (Photo Q). 

6

3.

Press the 

EXECUTION

button again and 

the message 

Calib PASS!

will appear on 

the readout (Photo R).

Photo P 

4.

Keep pressing the 

EXECUTION

button to 

scroll through the information for baskets 1 
and 2: 

+

Average will be shown,  

PASS

message 

appears.

+

The 

Diff = (%)

message will appear.


Photo R 

Photo Q 

+

The 

Diff PASS

or

Diff FAIL

message will appear.  If 

Diff FAIL

appears, then one of the two goals must 

be adjusted to within 5% of the reading of the other goal. 

+

The final message will indicate whether the court has met the criteria for Rebound Range and Rebound 

Differences.

+

A

Court PASS

or

Court FAIL

message will appear. 

5.

This concludes the test. 

6.

The last page of this booklet is a log sheet.  Make a copy of this sheet, and record the test results for your 
records.

7.

Make certain the unit is turned off.  Return it to the carrying case immediately. 


TEST RESULTS

Since NCAA (Div. I) specifies semi-annual testing (once before 
the season and once before the post-season), it is recommended 
that a dated log sheet be filed, to prove that proper testing and 
adjustments were performed.  See the 

TEST PROCEDURE LOG 

SHEET 

on the back of this booklet – make additional copies for 

future testing. 

The Fair-Court

®

test device should also be readily available during 

all conference/competition games, to test the equipment (rebound 
elasticity and rim height) should any complaint or protest arise in 
regard  to  equipment  not  meeting  NCAA  performance 
requirements.

Summary of Contents for FAIR-COURT

Page 1: ...N MAINTENANCE MANUAL FAIR COURT OFFICIAL BASKETBALL EQUIPMENT TESTING SYSTEM No ERTD2003NCAA SAVE THESE INSTRUCTIONS FOR FUTURE USE INST 00378 063 2011 PORTER ATHLETIC EQUIPMENT COMPANY ALL RIGHTS RES...

Page 2: ...y and a new battery should be installed immediately Battery life is 20 40 hours depending on battery quality and use When in execute mode Drop Weight uses the most electricity and will drain the batte...

Page 3: ......

Page 4: ...give the percentage of the energy absorption 13 Continue pressing the EXECUTION button to repeat for Calib Drop 2 and Calib Drop 3 After Drop 3 the average of the three drops will be shown on the rea...

Page 5: ...ure to avoid possible injuries or damage to the unit 3 Press the MODE button and the message Test Bskt 1 will appear on the readout Photo M 4 Press the EXECUTION button and the message Bskt 1 Drop 1 w...

Page 6: ...r goal The final message will indicate whether the court has met the criteria for Rebound Range and Rebound Differences A Court PASS or Court FAIL message will appear 5 This concludes the test 6 The l...

Page 7: ...PROBLEM Reading seem erroneous or unordinary SOLUTION If you are in a mode such as Calibration or Basket 1 or 2 you may cycle through and replace previous data SOLUTION If readings are still out of t...

Page 8: ...rage 2 Backstop No 1 Passes Fails 3 Backstop No 2 Passes Fails 4 Difference Passes Fails NCAA CALIBRATION REQUIREMENT The rebound elasticity of any basket ring support system shall be within a 35 perc...

Reviews: