3.10 DC Test (300V)
In this unit, users can make DC sample and calibrate DC voltage transmitter according to the rated
value.
U1-2 V
Out
%
U300 V
Out
%
DC TEST (300V)
Trans.
Range
OutType
Start
Udc In
RET
0
100
0
100
U1-2
4-20mA
1-Way
0.00000
DC In mA Err %
0.0000
U1-2 V
U300 V
0.000
0.000
Figure 3-9 DC 300V calibration interface
DC Test (300V) interface includes:
Right: users can set calibration parameters;
Left: Four-phase voltage amplitudes and phase output values are based on the users’
setting;
Lower left: DC measurement result is showed in DC ln area. User presses the second
button to select DC current measurement value (input IdcL) or DC voltage measurement
(input Udc). The transmitter calibration error is showed in Err area.