20
The
Dose Rate
and
Dose
buttons are used to switch the
detector between the gamma-radiation DeR measurement mode
(indication units are
µSv/h, mSv/h, Sv/h or µR/h, mR/h, R/h
),
and the De measurement mode (indication units are
µSv, mSv,
Sv
or
µR, mR, R
), the De accumulation time is displayed in days,
hours and minutes.
While in the DeR measurement mode press the
Reset
button
to reset accumulated DeR statistical data and to restart a new
accumulation. Press the button (in the upper right corner)
to open the settings window where the User can change DeR
thresholds, DeR measuring units, enable or disable the sound
alarm or restore the settings by default. The default DeR threshold
settings are:
DeR Threshold 1: 1.0
µ
Sv/h;
DeR Threshold 2: 10
µ
Sv/h.
While in the De measurement mode press the
Reset
button to
reset the accumulated dose and De accumulation time values.
Press The button (in the upper right corner) to open the settings
window where the User can change De thresholds, De measuring
units, enable or disable the sound alarm or restore the settings by
default. The default De threshold settings are:
De Threshold 1: 10 mSv;
De Threshold 2: 10 Sv.