Polimaster PM2010 Operating Manual Download Page 20

 

 

20 

6 STORAGE AND SHIPPING 

 

6.1

 

Devices are to be stored in the manufacture’s package at the air temperature from -

20 to 50 

C and relative humidity up to 95 % at temperature of +35 

C.  Storage duration should 

not exceed the average service life, i.e. 10 years. 

 
Devices  without  package  are  to  be  stored  at  the  air  temperature  from  10  to  35 

C  and 

relative humidity of 80 % at temperature of 25 

C. 

 

The  storage  place  should  be  free  of  dust,  vapors  of  strong  chemicals  and  other 

substances that may cause corrosion. 

 

6.2  Devices  in  package  may  be  shipped  by  any  kinds  of  enclosed  transport  at  the    air 

temperature from -20 to 50 

C.  

Devices  in  package  should  be  fastened  in  a  vehicle.  They  must  be  arranged  and 

fastened in transport so as to ensure their stable position and to avoid shocks to each other and 
the walls of a vehicle. 

 
When carried by see, devices in package should be placed in  hermetic plastic bags with 

silicagel. 

 
When carried by air, the devices in package should be placed in hermetic compartments.  

 

The environmental conditions during transportation should be within the limits as follow: 

-

 

air temperature from - 20 to 50 

C; 

-

 

relative humidity up to 100% at temperature 40 

C. 

 

 

Summary of Contents for PM2010

Page 1: ...ALARMING DEVICE FOR DETECTION OF GAMMA RADIATION AND CHEMICAL WEAPONS VAPORS PM2010 OPERATING MANUAL ...

Page 2: ...3 1 Preparation for use 13 3 2 Device operation 14 4 MAINTENANCE 16 5 TROUBLESHOOTING 19 6 STORAGE AND SHIPPING 20 7 UTILIZATION 21 8 WARRANTY 22 9 PACKING CERTIFICATE 23 10 CERTIFICATE OF PUTTING INTO SERVICE 24 11 ACCEPTANCE CERTIFICATE 25 12 WARRANTY CERTIFICATE 26 13 CLAIMS 27 APPENDIX A Copy of certificate toxic substance detection module tsm 28 APPENDIX B Beta radiation source copy of passpo...

Page 3: ...nance as well as some other information necessary for the proper operation of the device and full realization of its possibilities The name of the device in the documentation is written as follows Alarming device for detection of gamma radiation and chemical weapons vapors PM2010 Type 100345122 042 2004 During manufacturing of the device some changes may be introduced in its electrical scheme cons...

Page 4: ...ration history is stored in its non volatile memory and can be transferred to a personal computer PC via an infrared IR interface of data transmission using IrDA interface The device may be operated both indoor and outdoor by a wide range users involved in detecting of ionizing radiation sources and vapors of toxic substances The device may be used by staff of emergency services by border guards c...

Page 5: ...operating manual Appendix B Certificate of toxic substance module 1 piece included in operating manual Appendix A Battery Energizer L91BP 2 AA 1 2 pieces IR adapter ACT IR220L or IR 210B 2 1 piece Set of spare parts and tools SET O Filter 1 piece Rod 1 piece Screwdriver 1 piece Filtering tissue element 4 pieces Filtering element 4 pieces Bottle with filter filler 1 piece Coarse calico strip 0 4 m ...

Page 6: ...the set thresholds are exceeded and record in non violate memory the events of set DER threshold exceeding and history of DER registration in case of supply voltage changes from 2 5 to 3 2 V 2 3 9 Threshold of detection with a toxic substance module at contents higher than threshold values at normal climatic conditions NCC temperature 20 5 o C atmospheric pressure 101 3 5 4 17 3 kPa relative humid...

Page 7: ...5 mm repeated shocks with acceleration of 100 m s2 duration of shock pulse of 2 50 ms at shock rate of 60 180 shocks per minute 2 3 18 Device is resistant to RF electrical and magnetic fields under IEC 61000 4 3 1995 in the frequency range of 80 to 1000 MHz in the frequency range of 800 to 960 MHz and of 1 4 to 2 0 GHz generated by digital mobile phones 10 V m severity degree 3 30 V m severity deg...

Page 8: ...he registered concentration value of toxic substance vapors with the set threshold value of concentration level The detection modules are operated by a microprocessor controller with a built in analogue digital converter ADC The device is powered by two AA batteries The nominal battery voltage is 1 5 V 2 4 2 A block diagram of the device is presented in Figure 1 The device consists of gamma radiat...

Page 9: ... 2 with a cap 15 is located on the top cover of the device When the device is operated the cap is taken off the connecting pipe 2 and fixed adjacent to the connecting pipe at a plug 1 During the device calibration the filter 14 involved in the set of spare parts and tools SET O is screwed on the connecting pipe 2 The Red LED 4 Yellow LED 5 Green LED 6 ON OFF button 8 window 7 of the IR interface a...

Page 10: ... window with the device serial number 4 Yellow LED 5 Red LED 6 Green LED 7 input window of the IR interface 8 ON OFF button 9 GMM geometric center 10 cap of the output connecting pipe 11 housing with batteries 12 window for the audio alarming device 13 clip 14 filter 15 cap 16 filter cover Figure 2 ...

Page 11: ...itions While in this mode the device produces light and audible signals as follows Chemical alarm Red LED is flashing when the threshold value of the TSM is exceeded Radiation alarm Yellow LED is flashing when the preset DER threshold is exceeded The light alarms are accompanied by audible alarms if these are enabled in the mode of communication with PC The device provides for a possibility to rec...

Page 12: ...on the device The Radiation danger sign with the source type and its activity is marked on the bottom side of the device housing A seal that prevents an unauthorized opening of the device is placed in the faceplate of the device in the screw groove 2 4 6 Packing The device is put into a pasteboard box together with its documents and SET ...

Page 13: ...crew cap 15 on the connecting pipe at a plug 1 and put cap 10 of the output connecting pipe into SET O 3 1 3 6 When starting the operation after the long storage of the device it is required to clean the gas system and the ionizing chamber in according to clause 4 4 of the present Manual 3 1 3 7 Calibrate the device according to clause 3 2 1 2 After calibration completion the device is ready for o...

Page 14: ... to ensure detection thresholds perform calibration until both the R LED and the Y LED are off 3 2 2 Device operation in the mode of detection of exceeding the preset thresholds of the gamma radiation DER and concentrations of toxic substances vapors 3 2 2 1 While in the mode of detection of exceeding the preset thresholds of the gamma radiation DER and concentrations of toxic substances vapors th...

Page 15: ...solution of a monitor IrDA Connect the cable of the IR adapter to a communication port of the personal computer or use the PC with a built in IR adapter To install the application program use the CD with software included in the delivery kit The Help file describes how to run the application program To communicate with PC turn the device ON and place it for 15 s near the IR adapter at a distance o...

Page 16: ...2 After the beta radiation source service life indicated in source passport Appendix B is expired it should be replaced by a new one The radioactive source replacement should be performed by the specially trained personnel of the manufacturer personnel of A group 4 2 3 The utilization of used sources should follow a determinate order observing the requirements of handling radioactive materials 4 3...

Page 17: ...To blow out the gas system it is necessary to turn the device ON the filter cover open and hold it ON for one hour False chemical alarms may occur After an hour turn the device OFF and close the filter cover CAUTION When the maintenance of the gas system and ionizing chamber is performed alcohol drops and tissue fibers should not be left in the ionizing chamber and connecting pipe 4 5 Replacement ...

Page 18: ...SET O 6 washer 7 spring 8 nut 9 cover Figure 3 4 6 Device deactivation 4 6 1 The maintenance involves the external examination dust and dirt removal and decontamination in case when radioactive substances are fallen on the device housing Decontamination is carried out by wiping the device housing with a soft tissue moistens with ethyl alcohol ...

Page 19: ... of TSM Cap of the ionizing chamber is not removed Blocking of the gas system and ionizing chamber Imitator cover is not removed TSM module is failed Remove the cap Clean the gas system and ionizing chamber according to clause 4 4 Remove the cover Send device to the manufacturer 4 Red LED and Yellow LED are flashing in turn for more than 2 minutes One of the modules is failed Send device to the ma...

Page 20: ...ay cause corrosion 6 2 Devices in package may be shipped by any kinds of enclosed transport at the air temperature from 20 to 50 C Devices in package should be fastened in a vehicle They must be arranged and fastened in transport so as to ensure their stable position and to avoid shocks to each other and the walls of a vehicle When carried by see devices in package should be placed in hermetic pla...

Page 21: ...ices of the group 3 The utilization of the used source should be performed according to the requirements of regulations and safety measures for the radioactive source handling should be observed 7 2 Precious metal data contented in the device are not quoted as their weight is not higher than the values stated in the Sate Standard GOST 2 608 78 ...

Page 22: ...period is 6 months since the date of manufacturer Quality Control Department accepting the device 8 4 Warranty and after warranty repairs are performed by the manufacturer or by a company authorized by the manufacturer 8 5 Warranty does not cover the devices subjected to the user s service the device was tampered opened by the user with mechanical damages if the requirements of operation and stora...

Page 23: ...___ serial number packed ______________________________________________________________ name of manufacturer According to the requirements of design documentation __________________________ ________________________ ______________________ position signature family name ____________________________________ year month date stamp ...

Page 24: ...or detection of gamma radiation and chemical weapons vapors PM2010 Serial number ______________ is put into service ______________________ date of putting into service ___ _________________ _________________________ signature and name of the responsible person ...

Page 25: ... vapors PM2010 Serial number ___________________ is manufactured to meet the requirements of State Standards and the Design documentation and accepted for operation Head of Quality Control Department stamp place _____________________ ________________ signature name ___ ________________ _____________ ...

Page 26: ...0 81 Fax 375 17 260 23 56 E mail polimaster polimaster com E mail plant polimaster com Address 112 M Bogdanovich str Minsk 220040 Republic of Belarus Address 51 Skoriny str Minsk 220141 Republic of Belarus Date of sale ___ ________ _____ Seller ______________ signature Seller stamp _________________ The warranty after warranty repair is made ___ ____________ ______ The warranty period is prolonged...

Page 27: ...27 13 CLAIMS ...

Page 28: ...anophosphorus substances sarin soman V gases and arsenic containing substances lewisite in air A 3 DELIVERY KIT A 3 1 TSM delivery kit includes Toxic substance detection module 1 piece Quality certificate passport of radioactive source Appendix 2 of the present Manual 1 piece SET O set of spare parts and tools 1 set batch of TSM delivered to one address is supplied with 3 operating manuals per 10 ...

Page 29: ...by the source external radiation nA ______________ B 2 4 Relative error of the ionization current value at 0 95 confidence level 15 B 2 5 The radioactive contamination of the source external surface does not exceed the permissible value An activity of radioactive substances taken from the source surface by the smear method of determination of the radioactive contamination level is no more than 5nC...

Page 30: ...erature from 50 o C to 50 o C and a relative humidity up to 98 at 40 o C B 6 3 Do not remove seals and open package during the source transportation and storage B 6 4 Sources should not be stored in rooms provided with a closed ventilation system as well as in the same room with inflammable explosive substances and aggressive chemicals B 7 OPERATION GUIDELINES B 7 1 When operating the source requi...

Page 31: ...31 ...

Reviews: