8
- on exposure to magnetic field of 400
A/m strength
5%
- on exposure to radio frequency
electromagnetic
fields
of
30
V/m
strength
5%
13
LCD backlight at pressing LIGHT/SET
button
3-5 s
14
Instability of readings during 24 hour
continuous work, no more than
5%
15
Response time when DER value is
increased more than 10 times, from
initial low DER value to finite fixed high
DER value exceeding 10
Sv/h, no
more than
5 s
16
Response time when DER value is
decreased more than 10 times, from
initial high to finite fixed low DER value
exceeding 10 Sv/h, no more than
10 s
17
PC communication
- by a special program using IR
communication channel adapter
18
In the mode of data transmission to PC
the dosimeter provides the following
functions:
1) permission or prohibition the following
be displayed on LCD:
- DER or DE photon radiation measured
value indication;
- relative mean square error of average
value of measurement result with 0.95
probability (statistical error) in percents;
-
dosimeter number (“blind dosimeter”);
2) reading from PC into dosimeter and
storing in instrumen
t’s non-volatile
memory the following information:
- dosimeter serial number;
- history of DE and DER accumulation at
user set time periods;
- DE (DER) values when set thresholds
are exceeded, as well as time, date and
month, when set thresholds where
exceeded;
- DE and DER set thresholds values;
- service information;
3) following information is recorded from
PC to dosimeter:
- values of DE and DER thresholds
being set;
- current date and time for DE
accumulation history;
- operation modes.
19
In the search mode the dosimeter
P
М1621М,
РМ1621МА)
enables
following functions:
- detect and locate the photon radiation
sources on the basis of audio-, vibro-
and light (LED blinking) alarm signal
frequency changes;
- search mode on/off;
- search mode on/off state indication.
Summary of Contents for PM1621
Page 1: ...X RAY AND GAMMA RADIATION PERSONAL DOSIMETERS 1621 Models 1621 1621 1621 1621 OPERATING MANUAL...
Page 13: ...13 Figure 1 2 General overview of the dosimeter...
Page 14: ...14 Figure 1 3 Demounting the clip from 1621 and 1621...
Page 21: ...21 Figure 2 2 Select the 1621 1621 dosimeter operating indication mode...
Page 22: ...22 Figure 2 3 Setting the DE threshold values...
Page 23: ...23 Figure 2 4 Setting the DER threshold values...