Instructions for Use 2019-03-20
• Check operation before each patient. If the indicators in the display do not all appear normally, the device may not be
able to make an accurate measurement. In this case, stop using the device and have it repaired.
Checking the Function
⹅
Checking the Function with the Tester
Check device performance with the tester once a week.
1. Press the Power Switch to turn the device on.
2. Insert the tester into the probe cord jack.
Check that the meter indicates within ±3 bars away from (above
or below) 1.
* The meter may jump when the tester is inserted. If it does, wait
for about 1 second until the meter stabilizes and then check the
reading.
* If the reading is 4 or more bars away from 1, the device will not
make an accurate measurement. In this case, contact Poldent Co.
Ltd.
Must be
within this
range
Tester