Nitronox Plus®

Quick Start Guide

Instructing the Patient

The  Nitronox  Plus®  System  provides  a  50%:50%,  0-50%  or 
0-70% ratio of oxygen to nitrous oxide gas; the gas mixture is 
used to treat pain and anxiety during medical procedures.

Patient  must  hold  the  mouthpiece/mask  themselves.  If  they 
are  unable,  choose  alternate  method  of  pain  control.  Have 
patient hold the circuit hose and instruct patient to maintain a 
seal with mouthpiece or mask while breathing in and out.

Pain-relief  will  occur  within  30-90  seconds  of  breathing  through  the  mouthpiece  with  deep 
breaths.  The  anxiolytic  effect  of  gas  mixture  will  dissipate  within  3-5  minutes  of  the  last 
inhalation. To ensure safety, watch the patient for 10 minutes post-use.  

The following are examples of instructions to give the patient when they are about to use the 
Nitrous Oxide Delivery System for the first time.  

“Take the hose with the mouthpiece. When you feel pain/anxiety starting, or when instructed, hold 
the mouthpiece in your mouth and take slow deep breaths. You will hear the pain-relief gas being 
delivered. Keep the mouthpiece in your mouth as you breathe in and out. Within a few breaths you 
will start to feel more relaxed, and the pain will be lessened. You may also feel a little bit dizzy and 
your voice may sound different.” 

“If you feel nauseous, remove the mouthpiece and breathe deeply from the room air. The feeling will 
quickly go away.”  

“The  gas  mixture  will  not  completely  remove  the  pain,  but  should  make  it  less  intense  and  more 
tolerable.”  

“When the pain lessens, remove the mouthpiece from your mouth and breathe normally. Keep the 
mouthpiece ready so you can use it during your next pain/anxiety event during your procedure.”  

PLYMOUTH MEDICAL

135 PLYMOUTH ST, SUITE 410

 

BROOKLYN, NY

 

11201

TEL – 888.392.5076

FAX – 888.392.1455

WEB – 

PLYMOUTHMEDICAL.COM

COPYRIGHT © 2020 PLYMOUTH MEDICAL LLC. ALL RIGHTS RESERVED.

V3-01-22

Summary of Contents for Nitronox Plus

Page 1: ... PRESSURE GAUGE IS NOT WITHIN THE 30 40 PSI GREEN ZONE 4 Place a patient circuit onto the front inlet and connect mouthpiece or mask to blue tubing Ensure the magenta tubing is pointed away from patient provider s 5 Turn the N2O Key Lock to the ON position to start administering the gas mixture 6 Set the desired concentration of nitrous oxide by adjusting the N2O dial on the front of the device It...

Page 2: ... System for the first time Take the hose with the mouthpiece When you feel pain anxiety starting or when instructed hold the mouthpiece in your mouth and take slow deep breaths You will hear the pain relief gas being delivered Keep the mouthpiece in your mouth as you breathe in and out Within a few breaths you will start to feel more relaxed and the pain will be lessened You may also feel a little...

Page 3: ... green as there is still gas in the lines no need to bleed the lines 6 Take the patient circuit off the mixer front inlet and dispose of the circuit magenta mouthpiece mask if used 7 Wipe the mixer off with disinfectant sani cloth do not spray product directly on device 8 The system is now shut down which should be done at the end of each day 9 We recommend removing the keys and storing them in a ...

Page 4: ...ulator off the yoke and replace the cylinder Ensure the regulator still has the nylon washer around the gas inlet There is no need to undo or loosen any other connections 3 Remove any protective wrap from around the yoke of the new cylinder Remove any FULL tags from new cylinder once you begin using it 4 Place the regulator over the yoke of the new cylinder lining up the index pins and gas inlet s...

Page 5: ...s oxide gas and oxygen for procedures You will go through 3 oxygen tanks to every 1 nitrous tanks per below 11 The device is ready for use ALWAYS LISTEN FOR LEAKS AT THE CONNECTIONS HISSING SOUND IF A LEAK IS DETECTED TIGHTEN THE FITTING TO THE REGULATOR T BAR WHERE THE HOSE SCREWS INTO THE SIDE OF THE TANK DEVICE IF WHISTLE ALARM SOUNDS DISCONTINUE PATIENT USE IMMEDIATELY AND SHUT OFF GAS SUPPLY ...

Reviews: