Nitronox Plus®
Quick Start Guide
Quick Steps to Start the System
1.
Using the cylinder wrench attached to your unit’s basket, turn the
nitrous oxide tank yoke valve on
first
by slowly turning the cylinder
wrench counter-clockwise (left) to fully open the flow of the tank
until you cannot open it further. Next turn the oxygen tank on by
turning the cylinder wrench counter-clockwise (left) to fully open the
flow of the tank.
2.
Once both gas tanks are turned on, ensure the O
2
supply pressure
and N
2
O supply pressure are within the green bands on the supply
line pressure gauges (45-60 psi). DO NOT USE DEVICE IF
EITHER OF THE PRESSURE GAUGES ARE NOT WITHIN THE
45-60 PSI GREEN ZONE.
3. Next, ensure that the mixture pressure gauge (center) is within the
green zone (30-40 psi). DO NOT USE DEVICE IF MIXTURE
PRESSURE GAUGE IS NOT WITHIN THE 30-40 PSI GREEN ZONE.
4. Place a patient circuit onto the front inlet and connect mouthpiece
or mask to blue tubing. Ensure the magenta tubing is pointed away
from patient/provider(s).
5. Turn the N
2
O Key Lock to the ON position to start administering
the gas mixture.
6. Set the desired concentration of nitrous oxide by adjusting the “%
N
2
O” dial on the front of the device. It is recommended to start with
a low percentage of N
2
O (30%) and titrate upwards until the desired
effect on the patient is achieved. The 50/50 model is not adjustable
and will deliver a preset 50%N
2
O/O
2
mixture. If, at any time, delivery
of 100% O
2
is desired, turn the “%N
2
O” dial to 0%, then turn the
N
2
O Key Lock to the OFF position. This will close the internal nitrous
oxide valve and supply the patient with 100% O
2
.