Page 20/22
Emissions Test
Conformity
Electromagnetic environment - Instructions
RF Emission CISPR 11
Group 1
The radiation is low and causes no interferences with
electronic devices found nearby.
RF Emission CISPR 11
Category B
The “plusoptiX S16” can be used in all areas which
are connected to the public mains supply.
Harmonic Emissions
IEC 61000-3-2
Category A
Voltage fluctuations /
Flicker IEC 61000-3-3
Complies
Immunity Test
IEC 60601 test values
Conformity values
Electromagnetic
environment -
Instructions
ESD IEC 61000-4-2
± 8 kV contact
± 2, 4, 8, 15 kV air
± 8 kV contact
± 2, 4, 8, 15 kV air
The floor should be
made of wood, ceramic
or stone. If the floor is
covered with a
synthetic material, the
relative air humidity
should not be less than
30%.
Electrical fast
transient / burst
IEC 61000-4-4
± 2 kV
± 2 kV
Only plugs that are
usually available in
domestic or clinical
areas should be used.
Surge
IEC 61000-4-5
± 0.5 kV
± 1 kV
± 2 kV
± 0.5 kV
± 1 kV
± 2 kV
Only plugs that are
usually available in
domestic or clinical
areas should be used.
Voltage dips, short
interruptions on
power supply input
lines IEC 61000-4-
11
0% 0.5 cycles
@ 0°, 45°, 90°, 135°,
180°, 225°, 270°, 315°
0% 1 cycle
70% 25 cycles
0% 250
cycles
0% 0.5 cycles
@ 0°, 45°, 90°, 135°,
180°, 225°, 270°, 315°
0% 1 cycle
70% 25 cycles
0% 250 cycles
Only plugs that are
usually available in
domestic or clinical
areas should be used.
Power frequencies
and magnetic fields
IEC 61000-4-8
30 A/m, 50/60Hz
30 A/m, 50/60Hz
Magnetic fields should
not exceed the usual
areas.