32
Table 3 Declaration – electromagnetic immunity
The PlayMakar
®
Pro Strength & Recovery System
is intended for use in the electromagnetic environment specified below.
The customer or the user should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment guidance
Electrostatic dis-
charge (ESD) IEC
61000-4-2
±6 kV contact
±8 kV air
±8 kV contact
±15 kV air
Floors should be wood, concrete or ceram-
ic tile. If floors are covered with synthetic
material, the relative humidity should be at
least 5%.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power supply lines
±1 kV for input/output lines
±2 kV for power supply
lines
Mains power quality should be that of a
typical commercial or hospital
environment.
Surge
IEC 61000-4-5
±1 kV differential mode
±2 kV common mode
±1 kV differential mode Mains power quality should be that of a
typical commercial or hospital
environment.
Voltage dips,
short interrup-
tions and voltage
variations on
power supply
input lines
IEC 61000-4-11
0 % U
T
; 0 , 5 cycle
At 0o , 45o , 90o , 135o ,
180o , 225o , 270o and 315o
0 % U
T
; 1 cycle and 70 % U
T
; 25/30 cycle single phase:
at 0o
0 % U
T
; 0 , 5 cycle
At 0o , 45o , 90o , 135o ,
180o , 225o , 270o and
315o 0 % U
T
; 1 cycle
and 70 % U
T
; 25/30
cycle single phase: at
0o
Mains power quality should be that of a
typical commercial or hospital environment.
If the user of the EQUIPMENT or SYSTEM
requires continued operation during power
mains interruptions, it is recommended that
the EQUIPMENT or SYSTEM be powered
from an uninterruptible power supply or a
battery.
www.PlayMakar.com
Summary of Contents for PRO Series
Page 1: ...Instruction Manual Model PRO 1000 ...
Page 37: ...37 www PlayMakar com ...
Page 38: ...38 www PlayMakar com ...
Page 39: ...39 www PlayMakar com ...