background image

user's manual

10020003_8

EN

Summary of Contents for SingLED

Page 1: ...user s manual 10020003_8 EN...

Page 2: ......

Page 3: ...1 Switching operating light on and off 3 5 2 Adjusting brightness of operating light 4 5 3 Adjusting handle positions 5 5 4 Adjusting patient mirror position 6 5 5 Attaching USB device to operating l...

Page 4: ...right to make changes without prior notice COPYRIGHT PLANMECA Publication number 10020003 Revision 8 Released 11 April 2014 The manufacturer assembler and importer are responsible for the safety relia...

Page 5: ...e device system including the Planmeca SingLED operating light must fulfil the medical device EMC requirements of the Standard IEC 60601 1 2 2 SYMBOLS Alternating current Standard IEC 60417 Planmeca S...

Page 6: ...sses that block high energy visible light HEV light or limit the direct exposure to 10 minutes CAUTION Do not leave the operating light on unattended CAUTION Never splash water on the operating light...

Page 7: ...t beam may speed up the hardening of some filling materials Move the light so that the light beam is not directed to filling material or dim the light 5 1 Switching operating light on and off The on o...

Page 8: ...ar two signal tones one after 3 and one after 15 seconds This mode is not available when the operating light is installed to a Planmeca Compact or a Planmeca Sovereign dental unit NOTE If the operatin...

Page 9: ...to another manufacturer s dental unit the brightness adjustment can be done only from the operating light 5 3 Adjusting handle positions Remove the handle by pulling it carefully outwards 1 Rotate the...

Page 10: ...d e g when they come off too easily Use only handles supplied by Planmeca see picture below 5 4 Adjusting patient mirror position When you want to turn the mirror away from the patient turn the mirror...

Page 11: ...lanmeca SingLED operating light 7 USING OPERATING LIGHT User s Manual 5 5 Attaching USB device to operating light The operating light has a USB port to which you can attach a USB device SingLED_USB ep...

Page 12: ...ustproof and impervious to spray vapour All exposed surfaces should be periodically wiped clean with a non abrasive cleaning agent designed for cleaning dental units When cleaning the light assembly m...

Page 13: ...35 C Remove the handle by carefully pulling it outwards 6 2 Arm The arm should be periodically wiped clean with a non abrasive cleaning agent designed to be used with dental units Stronger agents can...

Page 14: ...malfunction in the normal operation the sensor indicator light starts to flash The reason for an erroneous operation may be one of the following Error in the infrared sensor Broken LED Too low supply...

Page 15: ...ponents containing hazardous materials must be disposed of in accordance with waste legislation and instructions issued by the environmental authorities The risks involved and the necessary precaution...

Page 16: ...SingLED operating light is connected to other equipment the current must be limited to max 2 A with e g a fuse Environmental requirements Operating conditions Temperatures 15 C to 35 C 59 F to 95 F R...

Page 17: ...e measured sample does not fulfil the following parts of ISO FDIS 9680 2007 E standard 5 2 7 Shadow the hard shadow is too large in vertical and horizontal direction 5 2 8 Colour rendering index Ra is...

Page 18: ...TECHNICAL SPECIFICATIONS 14 Planmeca SingLED operating light User s Manual Dimensions 2 4 2 4 115 248 166...

Page 19: ...Planmeca SingLED operating light 15 TECHNICAL SPECIFICATIONS User s Manual 1 1 4 9 0 9 0 0 6 3 217 174...

Page 20: ...light User s Manual Ceiling installation Top view Side vie Maximum room height 3000 mm First joint rotation angle with rotation limit of ceiling adapter Second joint rotation angle with rotation limi...

Page 21: ...IFICATIONS User s Manual Wall installation Side view Top view Recommended max height from floor 1670 mm Maximum reach of outermost point of operating light First joint rotation angle with rotation lim...

Page 22: ...eca SingLED operating light User s Manual Dental unit installation Side view Top view Maximum reach of outermost point of operating light Second joint rotation angle with rotation limit First joint ro...

Page 23: ...Planmeca SingLED operating light 19 TECHNICAL SPECIFICATIONS User s Manual Dimensions of the wall plate in mm...

Page 24: ...TECHNICAL SPECIFICATIONS 20 Planmeca SingLED operating light User s Manual Dimensions of the ceiling adapter in mm Dimensions of ceiling adapter cover in mm...

Page 25: ...Planmeca SingLED operating light 21 MAINTENANCE PARTS User s Manual 10 MAINTENANCE PARTS Part Part number Power supply MENB1030A2402F01 24V DC 100 240VAC 10017273 Handle 10020763...

Page 26: ......

Page 27: ......

Page 28: ...Planmeca Oy Asentajankatu 6 00880 Helsinki Finland tel 358 20 7795 500 fax 358 20 7795 555 sales planmeca com www planmeca com...

Reviews: