3.3.4 X-ray unit settings
NOTE
If 60 kV can be set on the X-ray unit, this setting is preferred.
The standard exposure values for F-speed film (e. g. Kodak Insight) can be
used.
Intraoral X-ray units for an adult patient
The following tables show the standard values for the exposure time and the
dose area product of an image plate for an adult patient.
DC emitter, 7 mA, Tube length 20 cm
Without X-ray field
limitation
X-ray field limitation
2x3
X-ray field limitation
3x4
60 kV
mGycm
2
60 kV
mGycm
2
60 kV
mGycm
2
Incisors
0.10 s
18.3
0.10 s
3.8
0.10 s
7.7
Premolars
0.14 s
25.6
0.14 s
5.4
0.14 s
10.8
Molars
0.19 s
34.8
0.19 s
7.3
0.19 s
14.7
Bite wing
0.21 s
38.4
0.21 s
8.1
0.20 s
15.5
DC emitter, 6 mA, Tube length 30 cm
Without X-ray field
limitation
X-ray field limitation
2x3
X-ray field limitation
3x4
60 kV
mGycm
2
60 kV
mGycm
2
60 kV
mGycm
2
Incisors
0.15 s
13.7
0.15 s
4.9
0.15 s
5.8
Premolars
0.20 s
18.2
0.20 s
3.8
0.20 s
7.7
Molars
0.27 s
24.6
0.27 s
5.2
0.27 s
10.4
Bite wing
0.28 s
25.5
0.29 s
5.6
0.28 s
10.8
Check and adjust the specific X-ray unit in accordance with the standard
values.