Chapter D - ADJUSTMENT & CALIBRATION
Planmeca Proline EC X-ray unit D-9
PANORAMIC X-RAY BEAM ADJUSTMENT
Technical Manual
Radiation beam too far to the right or left
WARNING
WARNING: DO NOT ADJUST THE POSITION OF THE PRIMARY SLOT MECHANISM
WHILE X-RAYS ARE BEING GENERATED. CHECK THE BEAM POSITION, ADJUST THE
POSITION WITH THE X-RAYS SWITCHED OFF AND THEN RECHECK THE POSITION. IF
THE BEAM IS STILL MISALIGNED REPEAT THE OPERATION.
NOTE
The ceph primary slots are also adjusted according to instructions given in this sec-
tion.
If the beam is too far to the left or right of the alignment rectangle it must be centered.
Figure 9
NOTE
A microprocessor controlled motor determines the position of each slot. If the beam is
readjusted to the left or right the new position must be reprogrammed into the memory.
Enter the service mode if you are not already in it.
Press the CTL-key.
Press and hold down the clock key until the number of the slot that you are checking appears
on the main display (SLO and the slot number) and starts to flash. You are now in the primary
slot calibration mode.
For example, if you have just checked the position of the beam from the panoramic slot, the
number 1 will appear on the display.
Adjust the position of the beam by pressing the patient positioning (mm) keys. The minus key
will move the slot mechanism, and beam to the left, and the plus key will move the slot mech-
anism, and beam to the right.
Check the position of the beam again and, if necessary, repeat the above alignment proce-
dure until the beam is within the alignment rectangle on the fluorescent screen.
CAUTION
Radiation is generated when the exposure button is pressed. Take adequate protection
measures. Keep the exposure time as short as possible.
mm
CTL
CTL
Beam too far to the right
Beam too far to the left
CTL
Summary of Contents for Proline EC ceph
Page 2: ......
Page 8: ...TOC 6 Planmeca Proline EC Panoramic X ray TABLE OF CONTENTS Technical Manual...
Page 205: ...Planmeca Proline EC X ray unit J 1 Chapter Technical Manual J WIRING DIAGRAM...
Page 206: ...Chapter J WIRING DIAGRAM J 2 Planmeca Proline EC X ray unit Technical Manual...
Page 208: ......
Page 209: ......