2.2 Eliminating sideways play of the foot control pedal
a)
Remove the foot control cover as described in section “Removing foot control cover” on page
H-1.
b)
Remove the foot control PCB as described in section “Replacing foot control PCB” on page
H-3.
c)
Loosen the M3x4 DIN 916 locking screw on the underside of the foot control and loosen the
locking plate half a turn.
d)
Eliminate the horizontal play of the pedal by slightly turning the adjusting rod. When making
the adjustment, ensure that the pedal centering spring touches the adjusting rod and the
centering plate on both sides as shown below.
e)
Tighten the locking plate but make sure that the adjusting rod does not move when doing this.
f)
Tighten the M3x4 DIN 916 locking screw on the underside of the foot control.
g)
Reassemble the foot control in reverse order.
h)
Calibrate the foot control as described in section “Calibrating foot control” on page H-15.
adjpedalpla
y.eps
The sideways play
must be eliminated.
M3x4 DIN 916
Adjusting rod
Locking plate
Adjpedal0199.eps
Centering plate
Adjusting rod
Pedal centering spring
WRONG
CORRECT
Summary of Contents for Compact I
Page 1: ...technical manual 10007462_23 EN...
Page 2: ......
Page 20: ...Chapter A GENERAL TECHNICAL DATA A 12 Planmeca Compact i EMC INFORMATION Technical Manual...
Page 212: ...Chapter H FOOT CONTROL H 18 Planmeca Compact i ADJUSTMENTS Technical Manual...
Page 213: ...Planmeca Compact i I 1 Chapter Technical Manual I DIAGRAMS...
Page 214: ...Chapter I DIAGRAMS I 2 Planmeca Compact i Technical Manual...
Page 216: ......
Page 218: ......
Page 220: ......
Page 222: ......
Page 224: ......
Page 226: ......
Page 228: ......
Page 230: ......
Page 232: ......
Page 234: ......
Page 236: ......
Page 238: ......
Page 240: ......
Page 242: ......
Page 243: ......