Chapter E - PREVENTIVE MAINTENANCE
Planmeca Compact i E-3
FOOT CONTROL
Technical Manual
5 FOOT CONTROL
Pedal
•
Operation check. If the pedal does not move smoothly, remove the cover and clean and oil
the pedal joint and all the surfaces which the pedal touches.
Chair control
•
Operation check.
6 CUSPIDOR
Water regulator
•
Change the filter.
•
Change the membrane.
•
Change and lubricate the plug.
•
Check for leaks.
Air regulator
•
Change the filter.
•
Change the membrane.
Fuses
•
Condition check
Bowl
•
Bowl filter condition. Replace if needed.
•
Glass filling/bowl rinsing is working.
•
Check the operation of the bowl microswitch.
Internal air and water pressures
•
Check the pressures in service mode 38. Adjust if needed, refer to 3.1 “Adjusting internal
water & air pressures” on page F-26.
Lift motor
•
Operation check.
•
Clean and grease the worm screw. If there is any rust on the worm screw or nut, or if they
are worn, replace them. Additionally, if you find any rust or damage, check for signs of
leaks and eliminate the cause of the leaks.
Summary of Contents for Compact I
Page 1: ...technical manual 10007462_23 EN...
Page 2: ......
Page 20: ...Chapter A GENERAL TECHNICAL DATA A 12 Planmeca Compact i EMC INFORMATION Technical Manual...
Page 212: ...Chapter H FOOT CONTROL H 18 Planmeca Compact i ADJUSTMENTS Technical Manual...
Page 213: ...Planmeca Compact i I 1 Chapter Technical Manual I DIAGRAMS...
Page 214: ...Chapter I DIAGRAMS I 2 Planmeca Compact i Technical Manual...
Page 216: ......
Page 218: ......
Page 220: ......
Page 222: ......
Page 224: ......
Page 226: ......
Page 228: ......
Page 230: ......
Page 232: ......
Page 234: ......
Page 236: ......
Page 238: ......
Page 240: ......
Page 242: ......
Page 243: ......