Chapter F - ADJUSTMENTS
F-26 Planmeca Compact i
PNEUMATICAL ADJUSTMENTS
Technical Manual
3 PNEUMATICAL ADJUSTMENTS
3.1 Adjusting internal water & air pressures
The filter/regulators control the internal water and air pressures. There are two ways to moni-
tor the pressures while adjusting the regulators: manually with a manometer (pressure gauge)
or using the unit’s own pressure sensors. The manometer is connected to the head of the
syringe handpiece and the pressures are adjusted similarly as explained for adjusting the
pressures using the unit’s own pressure sensors.
Air pressure
a)
Enter service mode 38.
Press the A-key
to enter the air pressure mode. The unit
automatically closes down the main water valve and releases the internal water pressure via
the flush valve into the drain. The internal air pressure is shown on the display. The pressure
is shown in bars (1bar = 100kPa = 14.5psi).
b)
Lift the cap of the air filter/regulator upwards and adjust the internal air pressure by turning the
regulator cap. Balance the pressure from the syringe air valve after every adjustment. The
factory preset value for the internal air pressure is 5.5 bar.
c)
Lock the filter/regulator to the adjusted position by pushing down the regulator cap.
Water pressure
a)
Enter service mode 38.
Press the B-key
to enter the water pressure mode. The unit
automatically closes down the main air valve and releases the internal air pressure via the
instrument multiplexer. The internal water pressure is shown on the display. The pressure is
shown in bars (1bar = 100kPa = 14.5psi).
b)
Lift the cap of the water filter/regulator upwards and adjust the internal water pressure by
turning the regulator cap. Balance the pressure from the syringe water valve after every
adjustment. The factory preset value for the internal water pressure is 2.8 bar.
c)
Lock the filter/regulator to the adjusted position by pushing down the regulator cap.
+
-
AdjBur
k_Inal_press
.eps
Lift to adjust,
push down to lock.
Factory preset values:
* water pressure 2.8 bar
* air pressure 5.5 bar
Summary of Contents for Compact I
Page 1: ...technical manual 10007462_23 EN...
Page 2: ......
Page 20: ...Chapter A GENERAL TECHNICAL DATA A 12 Planmeca Compact i EMC INFORMATION Technical Manual...
Page 212: ...Chapter H FOOT CONTROL H 18 Planmeca Compact i ADJUSTMENTS Technical Manual...
Page 213: ...Planmeca Compact i I 1 Chapter Technical Manual I DIAGRAMS...
Page 214: ...Chapter I DIAGRAMS I 2 Planmeca Compact i Technical Manual...
Page 216: ......
Page 218: ......
Page 220: ......
Page 222: ......
Page 224: ......
Page 226: ......
Page 228: ......
Page 230: ......
Page 232: ......
Page 234: ......
Page 236: ......
Page 238: ......
Page 240: ......
Page 242: ......
Page 243: ......