Note
. The Kryo550-16 systems are designed to be connected to an Installation Category
(over-voltage category) II mains supply as defined in IEC 664.
Fuses
There are no user replaceable fuses. The Kryo550-16 mains inlet is fitted with a circuit
breaker to IEC/EN 60934, UL 1077.
10 A
230 V~
15 A
115 V~
Nominal trip at +23 ºC
Freezer type
Internal fuses must only be replaced by suitably trained service personnel.
For continued protection against electric shock, fuses must only be replaced
with fuses of the same type and rating
The controller includes two internal fuses that should only be replaced by suitably trained
service personnel. These must only be replaced after any fault has been cleared.
F1
T 1A L 250V IEC 127
F2
T 5A L 250V IEC 127
Batteries
Batteries must only be replaced by suitably trained service personnel.
This product contains a lithium battery which is an explosion hazard if
short-circuited or charged. Wear safety goggles when handling and take care
not to short-circuit the battery when handling and ensure that the terminals
are not inadvertently connected to any other live terminals.
This product contains a rechargeable battery. Always ensure correct polarity
is maintained when connecting and never short circuit or overcharge. Always
replace with a rechargeable battery of the same type. NEVER connect a
non-rechargeable battery as it could explode in use.
The equipment contains two batteries; a rechargeable 8.4V NiMH PP3 battery and a Lithium
TCL AA battery. These must only be replaced by suitably trained personnel.
A-2 Specifications
Summary of Contents for Kryo550-16
Page 1: ...Kryo550 16 Operator s Manual...
Page 9: ...C 1 Safety Declaration Contents vii...
Page 10: ...viii Contents...
Page 18: ...1 8 Introduction to the Kryo550 16...
Page 24: ...2 6 Installing the Kryo550 16...
Page 30: ...3 6 Freezing Samples...
Page 34: ...4 4 Routine Maintenance and Troubleshooting...
Page 38: ...B 2 Alarm Output...
Page 41: ......
Page 42: ......