Piezoconcept TT2 Installation And Operation Manual Download Page 4

 

 

 

1

 

INTRODUCTION  

 

The  TT2  is  a  two-axis  PZT  actuated  rotational  stage  of  exceptional  resolution  and 
stability.  The  TT2  comes  complete  with  position  sensitive  detectors  for  closed  loop 
operation. 

 

 

ROTATION (mrad) (XY)  

5mrad  / 10mrad 

VOLTAGE RANGE (V)  

-5V to +150V  

RESONANT FREQUENCY  
UNLOADED (Hz)  

(X) 4 kHz / 2.2 kHz 
(Y) 2 kHz / 1.2 kHz 

CABLE LENGTH (ft)  

>1.5m  

CABLE CONNECTION  

DB-9  

 

1.1

 

Unpacking the TT2 

Before unpacking the TT2, read this entire operation manual, paying special attention to 
the following section on 

“Handling the 

TT2

. Check the contents of the package against 

the shipping list and notify PIEZOCONCEPT immediately if any items are missing. 
 

1.2

 

Handling the TT2. 

The TT2 is a high precision scientific instrument and therefore requires special handling 
in  order  to  ensure  proper  operation.  Mishandling  can  cause  permanent  damage.  To 
insure a long and useful life the following guidelines should be strictly followed.  
 

 

Never  insert  anything  into  the  EDM  grooves

.  The  EDM  grooves  are  the  cuts 

that form the flexure hinges and separate the moving portion of the stage from the 
stage frame. Severe damage may result if objects are inserted into these grooves.  

  Do  not  move  the  rotational  stage  by  pushing  on  it  with  your  hands  or  any  other 

object.  

  Avoid applying a torque between the moving stage and the frame.  

  Do not drop, treat roughly, or physically shock the rotational stage.  

  Do not lift by the cable.  

  The  surface  to  which  the  TT2  is  mounted  to  should  be  flat  and  clean.  Likewise, 

the bottom of the TT2 should be free of particles and dust before mounting.  

  Do not immerse in any liquid. If the  TT2 requires cleaning slightly dampen  a lint 

free cloth with iso-propanol or ethanol and lightly wipe the surface. Do not get any 
liquid or lint into the EDM grooves.  

  Never  disassemble  the  nanopositioning  stage,  there  are  no  serviceable  parts 

inside.  
 

Summary of Contents for TT2

Page 1: ...1 INSTALLATION AND OPERATION MANUAL TT2 NANOPOSITIONER PIEZOCONCEPT 15 Rue du Bocage 69008 LYON France ...

Page 2: ...om the PIEZOCONCEPT controller first set the command voltage to 0 0V then turn the AC power to the PIEZOCONCEPT controller off and finally wait one minute before disconnecting The TT2 has no user serviceable parts Only trained service personnel should perform service IMPORTANT All Technical Information recommendations and examples related to PIEZOCONCEPT Products made in this manual are based on i...

Page 3: ...cking the TT2 4 1 2 Handling the TT2 4 1 3 TT2 5 2 INSTALLATION 5 3 GROUND LOOPS 6 3 1 Prevention and identification of ground loops 6 4 OPERATING THE NANOPOSITIONING STAGE 6 4 1 Operating in closed loop mode 7 4 2 Care during operation 7 ...

Page 4: ...t damage To insure a long and useful life the following guidelines should be strictly followed Never insert anything into the EDM grooves The EDM grooves are the cuts that form the flexure hinges and separate the moving portion of the stage from the stage frame Severe damage may result if objects are inserted into these grooves Do not move the rotational stage by pushing on it with your hands or a...

Page 5: ...e surface to be mounted to is clean flat and free of burs Using a lint free cloth gently wipe the bottom or mounting surface of the TT2 to remove any particles or dust Always lift the TT2 by the Y axis bottom portion Never lift position attach or unattach the TT2 with the power applied IMPORTANT Check for ground loops Section 3 between the TT2 and the mounting surface To install use the M2 CHC hol...

Page 6: ...er ground to the mounting surface The stage is connected directly to the ground of the PIEZOCONCEPT s controller which in turn is connected to the ground of the AC power cord The PIEZOCONCEPT s controller enclosure is also at ground potential Attaching a grounding wire between any of the PIEZOCONCEPT s controller enclosure screws and the mounting surface may short circuit the ground loop In a few ...

Page 7: ...witch on The command voltage now controls the position of the nanopositioning stage Never disconnect the 9 Pin connectors with the power on Always set the command voltage to zero and turn the power off before disconnecting Allow 1 minute for the PZT actuators to discharge before disconnecting For more information see the PIEZOCONCEPT S CONTROLLER OPERATION MANUAL 4 2 Care during operation The TT2 ...

Reviews: