PIE PIECAL 820 Operating Instructions Manual Download Page 4

Page 5

Page 4

SOURCE

HI

SET

LO

MAX

READ

MIN

READ

 

EZ-CHECK™ SWITCH

SOURCE:

 Instantly output two preset settings by moving the 

EZ-CHECK™ switch to the 

“LO” 

position or 

“HI”

 position. For fast 

three point checks select the 

“SET”

 position. The PIECAL 820 

will remember the last 

“SET” 

value, even with the power off.

These values can easily be changed to suit the calibration requirements. 
The values stored in the HI and LO positions are also used for Auto 
Stepping.

READ:

 Slide the switch to the SET position. The PIECAL 820 will 

display the current reading from the sensor or device being measured.  
Slide the switch to MAX and the highest value measured since turn-on 
or reset will be displayed; slide the switch to MIN and the lowest value 
measured since turn-on or reset will be displayed.

 

SOURCE/OFF/READ Switch

Select “

SOURCE

” to output mA, V, T/C, Ohms, RTD or frequency.

Select “

READ

” to read mA, V, T/C, Ohms, RTD or frequency.

 

EZ-DIAL™ KNOB

SOURCE: Turn the knob to adjust the output level. Turn clockwise to 
increase the output, counter clockwise to decrease the output in one 
least significant digit steps at a time. Push down and turn the 
EZ-DIAL knob for faster dialing.
Press and hold the knob for two seconds to store desired 
EZ-Check™ HI/LO points in SOURCE mode. Continue to press and 
hold the knob for two more seconds to start the automatic ramping.

READ:  Press and hold to transfer the current temperature into 
the EZ-Check™ MIN/MAX points. This clears the MIN/MAX 
readings which will update as the input value changes.

SELECTING FUNCTIONS

The EZ-DIAL knob is used to setup the PIECAL 820 to match 
the instrument to be calibrated or signal to be measured. Each 
time you turn on the PIECAL 820  the LCD displays the following 
screen for about 1 second followed by operating in the function 
used the last time it was operated.

Double Click the EZ-DIAL knob to change the function of the 
calibrator and to select ranges, units and other user settings. 
Each function (mA, V, T/C, Ohms, RTD, Frequency) has two 
pages of menus. The first menu page has settings for the func-
tion and the second menu page has settings for STEPPING, 
AUTO OFF and BACKLIGHT. Settings are remembered even 
with the power off but are reset when the batteries are changed.

PIECAL 820

DOUBLE CLICK

EZ-DIAL KNOB

FOR CONFIGURATION

Operating Instructions

Basic Operation

Operating Instructions

Basic Operation

Summary of Contents for PIECAL 820

Page 1: ...Practical Instrument Electronics Please contact an authorized sales representative at 800 346 4620 or order at www Instrumentation com ...

Page 2: ...imulate 2 Wire Transmitters Power up transmitters loops Calibrate in temperature with 0 1 C F resolution Types J K T E R S B N G C D L J DIN U T DIN P II Pt 100 Ohm 3850 3902 3916 3926 1000 Ohm 3850 RTD Copper 10 50 Ohm Nickel 100 and 120 Ohm RTD Checkout flow and vibration systems Source read frequency from 0 to 2000 CPM Counts Per Minute 0 00 to 999 99 Hz 0 0 to 9999 9 Hz and 0 000 to 20 000 kHz...

Page 3: ...ged externally are available STORING HI and LO EZ CHECK Source Outputs Speed up your calibration by storing Span Zero output setting for instant recall with the EZ CHECK switch 1 Store your high SPAN output temperature by moving the EZ CHECK switch to the HI position and turning the EZ Dial knob until the desired output value is on the display Press and hold the EZ Dial knob until STORED appears t...

Page 4: ...igit steps at a time Push down and turn the EZ DIAL knob for faster dialing Press and hold the knob for two seconds to store desired EZ Check HI LO points in SOURCE mode Continue to press and hold the knob for two more seconds to start the automatic ramping READ Press and hold to transfer the current temperature into the EZ Check MIN MAX points This clears the MIN MAX readings which will update as...

Page 5: ...T FUNCTION T C UNITS C F T C TYPE J K T E R S B G C D L U P COLD JUNC ON EXIT FUNCTION RTD UNITS C RTD Pt 100 3850 RTD Types WIRE MODE 234W EXIT FUNCTION RTD UNITS C RTD Pt 100 3850 RTD Types WIRE MODE 3W 2W 4W RTD Types Pt 100 a 3850 a 3902 a 3916 a 3926 Pt 1000 a 3850 Cu 10 a 4274 Cu 50 a 4280 Ni 120 a 6720 Ni 110 a 5801 Source mA Simulate 2 Wire Transmitters Read mA Power Measure Transmitters S...

Page 6: ... change the STEPS and the STEP TIME settings These settings are remembered even with the power off EXIT MENU exits this menu immediately and saves any changes Menu will automatically exit after 15 seconds of inactivity AUTO OFF If AUTO OFF is ON the unit will turn off after 30 minutes of inactivity to save battery life If AUTO OFF is OFF the unit will stay on until the POWER SWITCH is moved to the...

Page 7: ...ion on the EZ CHECK switch The output is adjusted in 0 01 mA 0 1 increments by turning the knob Press and turn the knob for faster dialing with 1 00 mA 10 0 increments To Sensor Typical 2 Wire Transmitter Disconnected IN REF OUT Receiver Powers External 2 Wire Transmitter Power Supply 2 to 100 VDC SOURCE HI SET LO MAX READ MIN READ 2 Wire SIM mA 2 Wire SIM Percent of 4 to 20 mA Choose this functio...

Page 8: ... hold the knob to clear the MAX and MIN readings All milliamp connections Typical 2 Wire Transmitter IN REF OUT Transmitter Input Sensor Process Signal Simulated Input SOURCE HI SET LO MAX READ MIN READ Power Measure mA Power Measure Percent of 4 to 20 mA Choose this function to simultaneously supply power to a 2 Wire Transmitter while displaying the 4 to 20 mA output of the transmitter Move the p...

Page 9: ...ECK switch The output is adjusted in 0 01 mV or V increments by turning the knob Press and turn the knob for faster dialing with 1 00 mV or V increments Process Calibrator mA V TC Ω RTD Hz SOURCE HI SET LO OFF MAX READ MIN READ OVERLOAD Hz SYNC Double Click Menu Push Turn for Fast Dialing Push Hold to Store Step READ mV 8 20 Voltage Output Signal Controller Transmitter Power Supply Read mV V Choos...

Page 10: ...uble Click Menu Turn the knob to scroll through the settings and press the knob to make your selection Select T C for the FUNCTION F or C for the UNITS T C Type J K T E R S B N G C D L J DIN U T DIN or P Platinel II and COLD JUNC ON or OFF ON is the default Connect the PIECAL 820 to the inputs of the device being calibrated using the proper type of thermocouple wire via the miniature thermocouple ...

Page 11: ...he knob to clear the MAX and MIN readings Resistance Source Choose this function to provide a simulated resistance into any device that measures resistance Move the power switch to SOURCE then Double Click the EZ DIAL knob to get into the Double Click Menu Turn the knob to scroll through the settings and press the knob to make your selection Select OHMS for the FUNCTION 400Ω or 4000Ω for the RANGE...

Page 12: ...nput devices that measure RTD sensors Move the power switch to SOURCE then Double Click the EZ DIAL knob to get into the Double Click Menu Turn the knob to scroll through the settings and press the knob to make your selection Select RTD for the FUNCTION F or C for the UNITS and RTD Choose from one of Platinum 100Ω or 1000Ω Copper 10Ω or 50Ω Nickel 120Ω or 110Ω curves Note Pt 100Ω 3850 is the most ...

Page 13: ... and press the knob to make your selection Select FREQ for the FUNCTION and 20KHZ 10000HZ 1000HZ or 2000CPM for the RANGE Disconnect all input wires from the devices to be calibrated and connect the PIECAL 820 to the input of the device matching polarity The green HZ SYNC LED pulses in synch with the output pulses and may be used to calibrate optical pickups The output signal is a zero crossing sq...

Page 14: ...st misconnection Over voltage protection to 60 vrms rated for 30 seconds Red LED indicates OVERLOAD or out of range conditions Display High contrast graphic liquid crystal display with 0 315 8 0 mm high digits LED backlighting Read mA Ranges and Resolution 0 00 to 24 00 mA or 25 0 to 125 0 of 4 20 mA Accuracy 0 03 of Full Scale Voltage burden 2V at 24 mA Overload Current limit protection 25 mA nom...

Page 15: ...ance 1 Megohms Open TC Threshold Pulse 10K Ohms 5 μamp pulse for 300 milliseconds nominal RTD OHMS and Continuity Read Resistance Ranges 0 0 to 401 0 0 to 4010 Ohms Accuracy 0 03 of Full Scale 0 075 Ohms Excitation Current 1 0 mA to 401 Ohms 0 6 mA to 4010 Ohms nominal Continuity 0 0 to 401 0 Ohms Beeps from 0 0 to 100 0 Ohms RTD and OHMS Source Accuracy From 1 to 10 2 mA External Excitation Curre...

Page 16: ...es C Range C Degrees F Range F J 200 0 to 150 0 1 2 328 0 to 238 0 2 0 150 0 to 50 0 0 7 238 0 to 58 0 1 3 50 0 to 100 0 0 5 58 0 to 212 0 0 9 100 0 to 1200 0 0 4 212 0 to 2192 0 0 8 K 230 0 to 150 0 2 6 382 0 to 238 0 4 7 150 0 to 0 0 1 0 238 0 to 32 0 1 8 0 0 to 1100 0 0 6 32 0 to 2012 0 1 1 1100 0 to 1371 1 0 7 2012 0 to 2500 0 1 2 T 260 0 to 230 0 6 1 436 0 to 382 0 11 0 230 0 to 150 0 2 2 382...

Page 17: ... 0 to 1500 0 1 3 752 0 to 2732 0 2 4 1500 0 to 2000 0 2 3 2732 0 to 3632 0 3 0 2000 0 to 2320 0 2 6 3682 0 to 4208 0 4 6 D 1 1 to 200 0 1 8 30 0 to 392 0 3 2 200 0 to 1350 0 1 4 392 0 to 2462 0 2 6 1350 0 to 2000 0 1 9 1742 0 to 3632 0 3 4 2000 0 to 2320 0 2 6 3632 0 to 4208 0 4 7 P 200 0 to 0 0 2 1 328 0 to 32 0 3 8 0 0 to 150 0 0 8 32 0 to 302 0 1 5 150 0 to 1100 0 0 6 302 0 to 2012 0 1 1 1100 0...

Page 18: ...y Our equipment is warranted against defective material and workmanship excluding batteries for a period of three years from the date of shipment Claims under warranty can be made by returning the equipment prepaid to our factory The equipment will be repaired replaced or adjusted at our option The liability of Practical Instrument Electronics PIE is restricted to that given under our warranty No ...

Page 19: ...ated on equipment traceable to NIST and includes a of Calibration Test Data is available for an additional charge Practical Instrument Electronics recommends a calibration interval of one year Contact your local representative for recalibration and repair services Shelby Jones Co 8800 W Chester Pike Upper Darby PA 19082 610 446 6600 or Toll Free 800 346 4620 www Instrumentation com ...

Reviews: