Configuration

Configure the Calibrator

Move 

POWER  SWITCH  to 

“SOURCE”   or 

“READ” 

Setup
Double click the 

e

 DIAL KNOB at any time the 

unit is on and the following displays will appear 

for 15 seconds:

Turn  the 

e

  DIAL  KNOB  to  move  through  the 

menu.  Press  the 

e

  DIAL  KNOB  to  toggle 

between  OFF  and  ON  or  to  scroll  through  the 

settings. 

EXIT MENU  - exits this menu immediately and 

saves any changes. Menu will automatically exit 

after 15 seconds of inactivity (countdown timer is 

displayed).

MODEL 422  V#.##

DOUBLE CLICK

EZ-DIAL KNOB

FOR CONFIGURATION

>  EXIT      

      15

 

TEMP UNITS  

      °C

 

T/C      

       K

 

CJC      

     ON

 

BACKLIGHT   

     ON

>  EXIT      

      15

 

STEPS    

 

        3

 

STEP TIME    

        8

 

AUTO OFF    

     ON

TEMP  UNITS   -  pressing  the  knob  will  toggle 

between °C and °F.

T/C   -  pressing  the  knob  will  cycle  through T/C 

types  J, T, E, K, R, S, B, N, G, C, D, L (J-DIN), U 

(T-DIN) and Platinel II. 

CJC   -  Automatic  CJC  (Cold  Junction 

Compensation)  may  be  turned  on  or  off.  It  is 

recommended that CJC be left on (default). CJC 

should  only  be  turned  off  if  an  external  cold 

junction  compensator  or  ice  bath  is  used  with 

the PIECAL 422.

BACKLIGHT   -  If  BACKLIGHT  is  ON  the 

backlight will light all the time the unit is powered 

up. For maximum battery life turn the backlight 

off  when  using  the  calibrator  in  areas  with 

enough ambient light to read the display.

STEPS  - pressing the knob will cycle through 2, 

3, 5, 11 and 21 steps. The endpoints of the steps 

are based on the values stored in the 

HI and  LO  

EZ-CHECK outputs.

STEP  TIME   -  pressing  the  knob  will  cycle 

through 5, 6, 7, 8, 9, 10, 15, 20, 25, 30 and 60 

seconds. 

AUTO  OFF   -  If AUTO  OFF  is  ON,  the  unit  will 

turn  off  after  30  minutes  of  inactivity  to  save 

battery life. If AUTO OFF is OFF the unit will stay 

on until the POWER SWITCH is moved to the off 

position.

SHOW mV  - If SHOW mV is ON the mV value 

corresponding  to  the  sourced  or  measured 

temperature is displayed.

SHOW CJ  - If SHOW CJ is ON the cold junction 

temperature  measured  by  the  PIECAL  422  will 

be displayed.

Note:  All settings are remembered even with the 

power  off.  Removing  the  batteries  resets  the 

values to factory defaults.

>  EXIT      

      15

 

SHOW mV    

     ON

 

SHOW CJ    

     ON

Please contact an authorized sales representative at 800-346-4620 or order at www.Instrumentation.com

Summary of Contents for PIECAL 422

Page 1: ...and Platinel II Easily set any value quickly to within 0 1 with the adjustable digital potentiometer DIAL plus store any three temperatures for instant recall with the EZ CHECK switch Calibrate quickl...

Page 2: ...he HI LO temperature readings which will update as the temperature changes Double click the knob to get into the PIECAL 422 Configuration Mode Use configuration to select C or F T C Type Backlight On...

Page 3: ...cold junction compensator or ice bath is used with the PIECAL 422 BACKLIGHT If BACKLIGHT is ON the backlight will light all the time the unit is powered up For maximum battery life turn the backlight...

Page 4: ...any other device 2 Select READ with slide switch w 3 Place the EZ CHECK switch into the READ position 4 Connect a thermocouple probe matching the type of wire to sensor being measured with min iature...

Page 5: ...the e DIAL KNOB for 6 seconds the word STORE will appear on the display after 3 seconds and continue to press the DIAL KNOB until the word STEPPING appears on the display The word STEPPING will appear...

Page 6: ...to 328 0 0 7 Chromel Constantan Jacket Purple Red Purple 200 0 to 100 0 0 2 328 0 to 148 0 0 4 100 0 to 850 0 0 1 148 0 to 1562 0 0 2 850 0 to 1000 0 0 2 1562 0 to 1832 0 0 4 R 13 3 to 250 0 1 2 1 0...

Page 7: ...2320 0 0 9 3812 0 to 4208 0 1 6 D 1 0 to 50 0 0 6 30 0 to 122 0 1 1 W3 Re W25 Re Jacket White Red White Yellow 50 0 to 1400 0 0 4 122 0 to 2552 0 0 7 1400 0 to 1800 0 0 5 2552 0 to 3272 0 0 9 1800 0...

Page 8: ...60 inches Weight 12 1 ounces including boot batteries Batteries Four AA Alkaline 1 5V LR6 Battery Life 50 Hours Optional NiMh Rechargeable battery kit 120 VAC for North America Only charger four NiMh...