Configuration
Configure the Calibrator
Move
w
POWER SWITCH to
“SOURCE” or
“READ” .
Setup
Double click the
e
DIAL KNOB at any time the
unit is on and the following displays will appear
for 15 seconds:
Turn the
e
DIAL KNOB to move through the
menu. Press the
e
DIAL KNOB to toggle
between OFF and ON or to scroll through the
settings.
EXIT MENU - exits this menu immediately and
saves any changes. Menu will automatically exit
after 15 seconds of inactivity (countdown timer is
displayed).
MODEL 422 V#.##
DOUBLE CLICK
EZ-DIAL KNOB
FOR CONFIGURATION
> EXIT
15
TEMP UNITS
°C
T/C
K
CJC
ON
BACKLIGHT
ON
> EXIT
15
STEPS
3
STEP TIME
8
AUTO OFF
ON
TEMP UNITS - pressing the knob will toggle
between °C and °F.
T/C - pressing the knob will cycle through T/C
types J, T, E, K, R, S, B, N, G, C, D, L (J-DIN), U
(T-DIN) and Platinel II.
CJC - Automatic CJC (Cold Junction
Compensation) may be turned on or off. It is
recommended that CJC be left on (default). CJC
should only be turned off if an external cold
junction compensator or ice bath is used with
the PIECAL 422.
BACKLIGHT - If BACKLIGHT is ON the
backlight will light all the time the unit is powered
up. For maximum battery life turn the backlight
off when using the calibrator in areas with
enough ambient light to read the display.
STEPS - pressing the knob will cycle through 2,
3, 5, 11 and 21 steps. The endpoints of the steps
are based on the values stored in the
HI and LO
EZ-CHECK outputs.
STEP TIME - pressing the knob will cycle
through 5, 6, 7, 8, 9, 10, 15, 20, 25, 30 and 60
seconds.
AUTO OFF - If AUTO OFF is ON, the unit will
turn off after 30 minutes of inactivity to save
battery life. If AUTO OFF is OFF the unit will stay
on until the POWER SWITCH is moved to the off
position.
SHOW mV - If SHOW mV is ON the mV value
corresponding to the sourced or measured
temperature is displayed.
SHOW CJ - If SHOW CJ is ON the cold junction
temperature measured by the PIECAL 422 will
be displayed.
Note: All settings are remembered even with the
power off. Removing the batteries resets the
values to factory defaults.
> EXIT
15
SHOW mV
ON
SHOW CJ
ON