Configuration

Power on & MAIN manu

Move 

POWER SWITCH to 

“SOURCE”

 or 

“READ”

Setup

Double click the 

e

 DIAL KNOB at any time the unit is on and the following displays will appear for 

15 seconds:

Turn the 

e

 DIAL KNOB to move through the menu. Press the 

e

 DIAL KNOB to toggle between OFF 

and ON or to scroll through the settings. 

EXIT MENU

 - exits this menu immediately and saves any changes. Menu will automatically exit after 

15 seconds of inactivity (countdown timer is displayed).

FUNCTION

 - pressing the knob will toggle between RTD and OHMS.

UNITS/RANGE

 - pressing the knob will toggle between °C and °F for RTD or 400Ω and 2000Ω for 

OHMS.

RTD

 - pressing the knob will cycle through the various RTD (Pt, Cu & Ni) at different base resistances 

and alpha values.

DISPLAY OHMS 

- If DISPLAY OHMS is ON the resistance associated with the RTD temperature will 

appear in small digits on the display.

Note:

 All settings are remembered even with the power off. Removing the batteries resets the values 

to factory defaults.

DOUBLE CLICK

EZ-DIAL KNOB

FOR CONFIGURATION

 V#.##

MAIN

   

>EXIT (1/2)
FUNCTION  RTD  

OHMS

UNITS 

°C 

°F

RTD      Pt 100 

a

=3850  

[*RTD Types - See Read RTD]

Source RTD

MAIN

   

>EXIT (1/2)
FUNCTION  RTD  

OHMS

UNITS 

    °C 

°F

RTD       Pt 100 

a

=3850

 Pt 1000 

a

=3850, Pt 1000 

a

=3750,

Pt 

100 

a

=3902, Pt 100 

a

=3916, Pt 100 

a

=3926,

Cu 

10 

a

=4274, Cu 50 

a

=4280, Ni 120 

a

=6720

Read RTD

>EXIT  

(1/2)   

  FUNCTION     OHMS
  RANGE 

    

400Ω 

2000Ω

MAIN

>EXIT

(1/2) 

  FUNCTION  

OHMS

  RANGE   

400Ω 

2000Ω 

Source Ohms

Read Ohms

Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176

 TestEquipmentDepot.com 

Summary of Contents for Evolution piecal 211

Page 1: ...itch Calibrate quickly with automatic output stepping Choose between 2 3 5 11 and 21 steps to automatically increment the output in 100 50 25 10 or 5 of span Select the step time to match your system...

Page 2: ...to start the automatic stepping READ Press and hold to transfer the current temperature into the EZ Check HI LO points This clears the HI LO temperature readings which will update as the temperature...

Page 3: ...cycle through the various RTD Pt Cu Ni at different base resistances and alpha values DISPLAY OHMS If DISPLAY OHMS is ON the resistance associated with the RTD temperature will appear in small digits...

Page 4: ...d LO EZ CHECK in 25 increments 0 25 50 75 100 11 steps between the HI and LO EZ CHECK in 10 increments 0 10 20 80 90 100 21 steps between the HI and LO EZ CHECK in 5 increments 0 5 10 90 95 100 Note A...

Page 5: ...LO position and turn the e EZ Dial knob until the desired temperature is on the display Press and hold the EZ Dial knob until STORED appears to store the value Release the EZ Dial knob 3 Instantly ou...

Page 6: ...alibrator where the jacks are numbered for ease of use SOURCE Plug in the 2 3 or 4 wires to match the connection on the instrument being calibrated READ Plug in the 2 3 or 4 wires from the sensor and...

Page 7: ...the knob for faster dialing with 10 or 1 00 10 0 ohm increments The OVERLOAD indicator will light if excessive voltage or current is detected by the calibrator Calibrating RTD Instruments 3 2 2 1 THE...

Page 8: ...splay with the current temperature reading Move the EZ CHECK switch to MAX to see the highest temperature reading and to MIN to see the lowest temperature reading Press and hold the knob e to clear th...

Page 9: ...connection due to a loose connector corrosion or a broken wire Here is an example of the PIECAL 211 reading a sensor with all 4 wire connected Here is an example where connections are made to a 4 wir...

Page 10: ...460 0 to 720 0 720 0 to 850 0 0 2 0 3 0 4 0 5 328 0 to 338 0 338 0 to 860 0 866 0 to 1328 0 1328 0 to 1562 0 0 4 0 6 0 8 0 9 Pt 1000 Ohm DIN IEC JIS 1989 1 3850 0 00385 200 0 to 230 0 230 0 to 550 0 5...

Page 11: ...onds Display High contrast graphic liquid crystal display LED backlighting for use in low lit areas Read Excitation Current 0 9 mA to 401 Ohms 0 4 mA to 2010 Ohms nominal Normal Mode Rejection 50 60 H...

Page 12: ...ipment will be repaired replaced or adjusted at our option The liability of Practical Instrument Electronics PIE is restricted to that given under our warranty No responsibility is accepted for damage...

Reviews: