Easy to use

With the PIECAL 211 you can check & calibrate all

your RTD instruments and measure RTD Sensors. 

Automatic indication of connections on the display

for simple hookups.

Take it into the shop, plant or field

Carry it without worry - it comes protected with a

rubber boot and rugged, low profile switch. Easy to

operate even in the dark areas of the plant with the

backlit display.

• Calibrate directly in temperature (°C & °F)

Stop carrying around a decade box and RTD

resistance tables. The PIECAL 211 works with the

RTDs you use including Platinum 100 (alpha =

3850, 3902, 3926) & 1000 (alpha = 3850, 3750)

Ohm, Copper 10 & 50 Ohm, and Nickel 120 Ohm. 

Easily set any value quickly to within 0.1° with the

adjustable digital potentiometer “

DIAL

” plus store

any three temperatures for instant recall with the

EZ-CHECK™ switch.

• Calibrate quickly with automatic output stepping

Choose between 2, 3, 5, 11 and 21 steps to

automatically increment the output in 100%, 50%,

25%, 10% or 5% of span. Select the step time to

match your system from 5, 6, 7, 8, 9, 10, 15, 20, 25,

30 and 60 seconds.

• Compatible with all process instruments

Connect directly to the RTD inputs of smart

transmitters, PLCs, DCS and multichannel recorders

and verify their outputs or displays. Works with older instruments with fixed excitation

currents and newer multichannel instruments that switch the excitation current

between input channels.

• Measure RTD Sensors

Troubleshoot sensor connections and find broken wires with patented technology. Connect your

two, three or four wire RTDs and the PIECAL 211 automatically detects the connections and

measures the RTD in degrees C or F. 

PIECAL 211

Automated RTD Calibrator

 

Operating Instructions

PIECAL 211

Automated RTD Calibrator

 

Operating Instructions

Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176

 TestEquipmentDepot.com 

Summary of Contents for Evolution piecal 211

Page 1: ...itch Calibrate quickly with automatic output stepping Choose between 2 3 5 11 and 21 steps to automatically increment the output in 100 50 25 10 or 5 of span Select the step time to match your system...

Page 2: ...to start the automatic stepping READ Press and hold to transfer the current temperature into the EZ Check HI LO points This clears the HI LO temperature readings which will update as the temperature...

Page 3: ...cycle through the various RTD Pt Cu Ni at different base resistances and alpha values DISPLAY OHMS If DISPLAY OHMS is ON the resistance associated with the RTD temperature will appear in small digits...

Page 4: ...d LO EZ CHECK in 25 increments 0 25 50 75 100 11 steps between the HI and LO EZ CHECK in 10 increments 0 10 20 80 90 100 21 steps between the HI and LO EZ CHECK in 5 increments 0 5 10 90 95 100 Note A...

Page 5: ...LO position and turn the e EZ Dial knob until the desired temperature is on the display Press and hold the EZ Dial knob until STORED appears to store the value Release the EZ Dial knob 3 Instantly ou...

Page 6: ...alibrator where the jacks are numbered for ease of use SOURCE Plug in the 2 3 or 4 wires to match the connection on the instrument being calibrated READ Plug in the 2 3 or 4 wires from the sensor and...

Page 7: ...the knob for faster dialing with 10 or 1 00 10 0 ohm increments The OVERLOAD indicator will light if excessive voltage or current is detected by the calibrator Calibrating RTD Instruments 3 2 2 1 THE...

Page 8: ...splay with the current temperature reading Move the EZ CHECK switch to MAX to see the highest temperature reading and to MIN to see the lowest temperature reading Press and hold the knob e to clear th...

Page 9: ...connection due to a loose connector corrosion or a broken wire Here is an example of the PIECAL 211 reading a sensor with all 4 wire connected Here is an example where connections are made to a 4 wir...

Page 10: ...460 0 to 720 0 720 0 to 850 0 0 2 0 3 0 4 0 5 328 0 to 338 0 338 0 to 860 0 866 0 to 1328 0 1328 0 to 1562 0 0 4 0 6 0 8 0 9 Pt 1000 Ohm DIN IEC JIS 1989 1 3850 0 00385 200 0 to 230 0 230 0 to 550 0 5...

Page 11: ...onds Display High contrast graphic liquid crystal display LED backlighting for use in low lit areas Read Excitation Current 0 9 mA to 401 Ohms 0 4 mA to 2010 Ohms nominal Normal Mode Rejection 50 60 H...

Page 12: ...ipment will be repaired replaced or adjusted at our option The liability of Practical Instrument Electronics PIE is restricted to that given under our warranty No responsibility is accepted for damage...