42
[Step 5]
Slide the metal cover out of the chassis as shown in Figure 4-5 and
complete the cover removal.
Figure 4-5
4.3
Main Board Removal
Follow the steps below to remove the main board. Of course, you must
complete the metal cover removal first.
[Step 1]
Remove the scanner card and GPIB card. If there are no scanner card and
GPIB card installed in your instrument, please skip this step.
Remove two fasteners of GPIB card and two captive screws that secure
the scanner card on the rear panel as shown in Figure 4-6. Pull out the
scanner card gently.
Summary of Contents for M3500A
Page 1: ...PICOTEST M3500A 6 5 Digit Digital Multimeter Service Manual Printed date 11 2011 ...
Page 52: ...52 4 5 1 Front Panel Assembly ...
Page 53: ...53 4 5 2 Chassis and Transformer Assembly ...
Page 54: ...54 4 5 3 Main Board Assembly ...
Page 55: ...55 ...
Page 56: ...56 4 5 4 Chassis Assembly ...
Page 57: ...57 ...
Page 82: ...82 5 3 Layout Drawings Main board PCB components layout top layer Figure 5 1 ...
Page 83: ...83 Main board PCB components layout bottom layer Figure 5 2 ...