USB TC-08 Single-Channel Terminal Board User’s Guide
Pico Technology
2
DO167-1
1
Safety notices
You MUST observe the following safety notices to prevent damage to
equipment and personal injury.
DO NOT do not connect the terminal board to any voltage source exceeding the
maximum input range printed on the board.
DO NOT connect the terminal board to a mains (line voltage) electrical supply. The
high voltage will damage or destroy the equipment and may cause serious or fatal
injury.
2
Overview
2.1
Introduction
The USB TC-08 Single-Channel Terminal Board is an accessory for the Pico
Technology USB TC-08 8-Channel Thermocouple Data Logger. The screw terminals
allow wires to be attached to the data logger without soldering and enable the USB
TC-08 to measure voltages from 0 to + 5 V or 4-20 mA loop currents.
The terminal board is designed for use with the USB TC-08 and is not guaranteed to
work with other thermocouple data loggers.
2.2
Specifications
Dimensions
57 x 27 x 14 mm (approx. 2.3 x 1.1 x 0.6 in.)
Weight
12 g nominal (approx. 0.5 oz)
Terminal wire size
1.5 mm
2
solid, 1.0 mm
2
stranded, 16-26 AWG
2.3
Inputs and switch
Name Function
4-20/5V
Input for the positive side of a 4-20 mA loop or 0 to
5 V signal
500mV
Input for the positive side of a 0 to 500 mV signal
50mV
Input for the positive side of a 0 to 50 mV signal
COM
Input for the negative side of any voltage or current
signal.
4-20 mA
Set switch to “ON” for a 4-20 mA loop signal.
Set switch to “OFF” for all voltage signals.