Pico Technology   

TA041 User’s Manual 

 

 

DO141-4 

© Copyright Pico Technology 2008–2012 

Avoid exposed circuitry 

To avoid injury, remove jewelry such as rings, watches and other metallic objects.   

Do not touch exposed connections and components when power is present. 

Use proper power source 

To ensure proper functioning of this probe, use four AA cells, a 6 V DC 60 mA 

mains adaptor, a 9 V DC 40 mA mains adaptor, or power leads.   

Do not operate the probe if it is damaged 

If you suspect there is damage to this probe, have it inspected by qualified service 

personnel.   

3.

 

Description 

By enabling conventional oscilloscopes to display and measure in-circuit 

waveforms that are referenced to high common-mode voltages, this differential 

probe extends the measurement capability of oscilloscopes to electronic power 

converters, inverters, motor speed controls, switch-mode power supplies and 

many other applications. 

4.

 

Installation 

Simply plug in the BNC output connector to the vertical input of a general-purpose 

oscilloscope or other measurement instrument, and connect the auxiliary 
grounding terminal to a proper ground.    The measurement instrument must have 

a ground reference. 

Install four AA cells or connect an appropriate power source to this probe. 

 

Select the proper attenuation ratio.    When measuring signals below 70 V, switch 

the attenuation ratio to 1:10 in order to get higher resolution and less noise.   

Otherwise, set the attenuation to 1:100. 
 

 

WARNING 

To  protect  against  electric  shock,  use  only  the 

accessories  designed  for  use  with  this  differential 

probe. 

 

Using the appropriate probe accessories, connect the input to the circuits under 

measurement. 
 

 

CAUTION 

This  probe  is  designed  for  carrying  out  differential 

measurements  between  two  points  on  the  circuit 

under  test.    It  is  not  intended  for  electrically 

insulating the circuit under test and the measuring 

instrument. 

Summary of Contents for TA041

Page 1: ...bols Terms appearing in this manual WARNING Warning statements identify conditions or practices that could result in injury or death CAUTION Caution statements identify conditions or practices that co...

Page 2: ...ou must verify that the probe is securely grounded before connecting the probe input leads Some types of measuring instrument such as a USB oscilloscope connected to a laptop are unlikely to be ground...

Page 3: ...ed controls switch mode power supplies and many other applications 4 Installation Simply plug in the BNC output connector to the vertical input of a general purpose oscilloscope or other measurement i...

Page 4: ...s The input leads of the differential probe connect to sprung hooks that come with the probe c Sprung Hooks The sprung hooks are connected safely to test points in the circuits under test 6 Power lead...

Page 5: ...6 dB 50 Hz 66 dB 20 kHz Ambient Operating Temperature 10 to 40 C Ambient Storage Temperature 30 to 70 C Ambient Operating Humidity 25 to 85 RH Ambient Storage Humidity 25 to 85 RH Power Requirements S...

Page 6: ...logy 6 Copyright Pico Technology 2008 2012 DO141 4 8 Derating curve The derating curve for absolute maximum input voltage is as follows 1000 1000 700 150 100 1000 10000 1 E 06 1 E 07 1 E 08 Voltage V...

Page 7: ...ainty in the test setup These are 1 The stated DC accuracy of the probe under test 2 2 Any DC offset or noise in the probe output The values in this manual are typical so to find the DC offset for a g...