Introduction
6
Copyright © 2012–2015 Pico Technology Limited. All rights reserved.
ps3000d.en r2
1.1.3
Grounding
WARNING
The oscilloscope's ground connection through the USB cable is for
measurement purposes only. The oscilloscope does not have a protective
safety ground.
WARNING
Never connect the ground input (chassis) to any electrical power source. To
prevent personal injury or death, use a voltmeter to check that there is no
significant AC or DC voltage between the oscilloscope ground and the point
to which you intend to connect it.
CAUTION
Applying a voltage to the ground input is likely to cause permanent damage
to the oscilloscope, the attached computer, and other equipment.
CAUTION
To prevent measurement errors caused by poor grounding, always use the
high-quality USB cable supplied with the oscilloscope.
1.1.4
External connections
WARNING
To prevent injury or death, only use the power adaptor supplied with the
product, where applicable. This is approved for the voltage and plug
configuration in your country.
PicoScope
model
USB connection
Ext DC power input
USB 2.0*
USB 3.0
Voltage (V)
Current
(A pk)
Total
power (W)
3203D
3204D
3205D
3206D
-
-
-
3403D
3404D
3405D
3406D
5 V
1.2 A
6.0 W
3203D MSO
3204D MSO
3205D MSO
3206D MSO
-
-
-
3403D MSO
3404D MSO
3405D MSO
3406D MSO
5 V
1.3 A
6.5 W
*
Can be powered from a single USB 2.0 port, provided the port supplies the
required power. The PicoScope software will notify you if the port is not suitable.
ООО "Техэнком" Контрольно-измерительные приборы и оборудование www.tehencom.com