PI Housed PICMA P-820 Short Instructions Download Page 4

M O T I O N   |   P O S I T I O N I N G

PZ303EK 1.2.0 11/2020. 2.0

Screwing on the Piezo Actuator

Tools and Accessories

 

Electrically conductive surface

 

Electrically conductive screw

 

Open-end wrench

 

Screwdriver or wrench

Requirements

 

Piezo actuator with stranded wires: The 

piezo actuator is discharged and short- 

circuited.

 

The surface has a mounting hole for fixing 

the piezo actuator.

 

The surface is connected to a protective 

earth conductor with a cross section of 

≥0.75 mm².

Screwing on the Piezo Actuator

1.  Secure the base of the piezo actuator with 

the open-end wrench.

2.  Fix the piezo actuator to the surface by 

tightening the screw (see manual for maxi-

mum torque).

Gluing the Piezo Actuator

Tools and Accessories

 

Electrically conductive surface

 

Electrically conductive adhesive (e.g., 

silver-filled epoxy resin)

Requirements

 

Piezo actuator with stranded wires: The 

piezo actuator is discharged and short- 

circuited.

 

The glued surfaces were prepared accor-

ding to the specifications of the adhesive 

manufacturer.

 

The surface is connected to a protective 

earth conductor with a cross section of 

≥0.75 mm².

Gluing the Piezo Actuator

1.  Apply the adhesive according to manufac-

turer's instructions.

2.  Spread the glue over the entire contact 

surface when gluing the piezo actuator.

Checking the Connection of the Protec

-

tive Earth Conductor

 

X

Make sure that the contact resistance at 

all protective earth connections is <0.1 Ω 

at 25 A.

 

X

If ground loops occur, contact the PI custo-

mer service department.

Screwing the Load

Tools and Accessories

 

Screw or nut

 

Open-end wrench

 

Screwdriver or wrench

Requirements

 

Piezo actuator with stranded wires: The 

piezo actuator is discharged and short- 

circuited.

Screwing the Load

1.  Secure the tip with the open-end wrench.

2.  Tighten the screw or nut (see manual for 

maximum torque).

Gluing the Load

Tools and Accessories

 

Adhesive (e.g., cold-hardening epoxy resin)

Requirements

 

Piezo actuator with stranded wires: The 

piezo actuator is discharged and short- 

circuited.

 

The glued surfaces were prepared accor-

ding to the specifications of the adhesive 

manufacturer.

Gluing the Load

1.  Apply the adhesive according to manufac-

turer's instructions.

2.  Spread the adhesive over the entire contact 

surface when gluing the load to the tip.

The instructions required for startup and 

operation are in the manual for the electronics 

used.

The piezo 

actuator must be discharged in the following 

cases:

 

Before short-circuiting

 

When not in use

 

Before reconnecting when the connecting 

cable was accidentally pulled out of the 

electronics during operation

 

Before assembly work

Starting and Operating the 

Piezo Actuator

Discharging and Short-

Circuiting the Piezo 

Actuator

Reviews: