0x79 Distance For Reference Search (Phys. Unit)
Used in: Reference move procedure (40)
Description:
Maximum distance for motion to the index pulse.
Data type
FLOAT
Command level
0
Item type
Axis
Source of data
Positioner database
0x7A Use Hard Stops For Referencing?
Used in: Reference point definition options (40)
Description:
Should the hard stops be used for reference moves?
Data type
INT
Command level
0
Item type
Axis
Source of data
Positioner database
Possible values
0
Do not use hard stops for reference moves
1
Use hard stops for reference moves
0x94 Notch Filter Frequency 1 (Hz)
Used in: Settings for the Notch Filter (35)
Description:
Frequency of the first notch filter.
Data type
FLOAT
Command level
0
Item type
Axis
Source of data
Positioner database
Possible values
40 to 20,000 [Hz]
10 PARAMETER REFERENCE
PZ285EN
‒
1/30/2019