DRT Set Data Recorder Trigger Source
Used in: Setting the trigger for recording (74), Starting the Recording (74)
Description:
Defines a trigger source for the specified data recorder table.
Long description
Irrespective of the trigger option set, the data recording is always
triggered when a step response measurement is carried out with
Trigger options:
<Trigger-
Source>
Description
0
default setting
Data recording is triggered with
<Value> is a dummy.
1
any command changing target position
Data recording is triggered by commands that in-
fluence the target position (e.g.,
, MVE).
<Value> is a dummy.
2
next command
Resets the trigger afterwards.
<Value> is a dummy.
3
external trigger
Data recording is triggered with a digital output
line.
<Value> is the ID of the digital output line.
6
any command changing target position
Corresponds to <TriggerSource> = 1, resets the
trigger afterwards.
<Value> is a dummy.
7
SMO command
Data recording is triggered with
, re-
sets the trigger afterwards.
<Value> is a dummy.
to query a list of all available recording and trig-
ger options.
Format:
DRT␣<RecTableID>␣<TriggerSource>␣<Value>
Arguments:
<RecTableID>
Data recorder table ID
<RecTableID> = 0 defines the trigger source for all
available data recorder tables.
<TriggerSource>
ID of the trigger source
<Value>
Depends on the trigger source
9 COMMAND REFERENCE
PZ285EN
‒
1/30/2019
129