6 Start-Up
E-836.1G Piezo Amplifier
PZ250E
Version: 1.0.1
17
In this Chapter
General Notes on Start-Up .......................................................................................... 17
Switching on the E-836.1G .......................................................................................... 18
Executing Motions ....................................................................................................... 19
6.1 General Notes on Start-Up
DANGER
Risk of electric shock if the protective earth conductor is not connected!
If a protective earth conductor is not or not properly connected, dangerous touch
voltages can occur on the E-836.1G in the case of malfunction or failure of the
system. If touch voltages exist, touching the E-836.1G can result in serious injury or
death from electric shock.
Connect the E-836.1G to a protective earth conductor (p. 13) before start-up.
Do
not
remove the protective earth conductor during operation.
If the protective earth conductor has to be removed temporarily (e. g. in the case
of modifications), reconnect the E-836.1G to the protective earth conductor
before starting it up again.
NOTICE
Reduced lifetime of the piezo ceramic due to permanently high voltage!
The permanent application of a high static voltage to the piezo actuators in the stage
reduces the lifetime of the piezo ceramic.
If the E-836.1G is not used, but is to remain switched on:
Set the control signal (sum of Control In and DC Offset) to 0 V.
6