MIC-200 Series / 1720

 

  Focus the image with the S40x lens. 

  Rotate  the  revolving  nosepiece  until  the  S100x  objective 

almost (not quite) clicks into place. 

  Drop a small drop of immersion oil (not part of the scope 

of  delivery,  please  order  separately)  on  the  center  of  the 
specimen. 

  Then let the S100x objective snap into place. 

  The front lens is now in contact with the immersion oil. 

  View  the  specimen  through  the  eyepiece  and  focus  the 

image with the fine adjustment knob. 

  The distance between objective lens and specimen is on-

ly 0.24mm! 

  If  small  bubbles  are  visible,  turn  the  S100x  objective  a 

few times to the left and right so that the front lens of the 
objective  moves  in  the  oil.  This  will  make  the  bubbles 
disappear. 

  Then turn the stage  with the fine adjustment knob down-

wards until the front lens does not touch the oil anymore.  

  Always clean the front lens of the S100x lens with a piece 

of lens paper moistened with a drop of xylene or alcohol. 

  Now also clean the micro slide. 

The S100x objective can also be used without immersion oil, 
i.e. dry. However, please note that in this case the resolution 
is  much  lower.  Water  may  provide  somewhat  better  resolu-
tion than using it dry. 

 
Caution! 

 

Never drop the xylene or alcohol directly onto the ob-
jective lens. This could cause xylene or alcohol to en-
ter the lens and dissolve the lens fixation!

 

 

Avoid oil contact with the other lenses! 

6  MAINTENANCE AND CLEANING 

Always cover the microscope with a dust cover after use. Al-
ways  leave  the  eyepieces  and  objective  lenses  mounted  so 
that no dust can enter the inside parts of the microscope. 
 

Cleaning the optical components 

If the eyepiece lenses or the front lenses of the objectives are 
dirty,  you  can  clean  them  by  wiping  a  piece  of  lens  paper 
over the lens surface in circular motions. If this does not help, 
use a drop of xylene or alcohol on the lens paper. Never put 
xylene or alcohol directly on the lens! 
If  dirt  is  clearly  visible  in  the  field  of  view,  it  is  on  the  lowest 
lens of the eyepiece. Remove the eyepiece from the tube and 
clean the outside of the lens. 
If  dust  is  still  visible,  check  whether  the  dust  is  in  the  eye-
piece by rotating it. If so, gently remove the bottom lens from 
the eyepiece and clean it. 
It is not necessary - and not recommended - to clean the lens 
surfaces  inside  the  objectives.  Sometimes  the  dust  can  be 
removed  by  high-pressure  treatment  with  air.  Otherwise, 
there will be no dust inside the objectives until the objectives 
are removed from the revolving nosepiece. 
The  most  important  step  to  a  long  lasting  clean  microscope 
optics  is  to  use  the  dust  cover  when  you  are  not  using  your 
microscope! 
 

Caution!  Cleaning  cloths  containing  plastic  fibers  can 
damage the coating of the lenses! 

 

Maintenance of the stand and mechanical components 

Simply remove dust with a brush. If the stand is very dirty, the 
surface can be treated with a careful cleaning agent. 
All moving parts such as the adjustment of the mechanical or 
the  coarse  and  fine  focusing  mechanism  contain  ball  bear-
ings  which  are  not  sensitive  to  dust.  The  bearings  can  be 

relubricated with sewing machine oil if necessary. 
 

Change the lamp and the fuse 

  Switch off the power and pull the mains plug. 

  Now loosen the lower cover screw of the chassis and re-

move the cover to replace the bulb and fuse according to 
the requirements of the different models. 

  After replacing the bulb and fuse, the cover should be re-

placed and the screws tightened. 

7  TECHNICAL DATA 

  Infinity  optical system 

  Siedentopf  viewing  head,  tubes:  30°  inclined  (binocu-

lar/trinocular), 360° rotatable 

  WF10x/18mm eyepieces for spectacle wearers 

  Eyepieces  of  the  monocular  models  secured  against  re-

moval 

  Quintuple nosepiece with click-stop 

  Achromatic objectives 4x NA 0.13, 10x NA 0.30, S40x NA 

0.70, and S100x NA 1.25 (oil)  

  Condenser: Abbe N.A. 1.25 with iris diaphragm 

  Models with mechanical stage: area 141 x 131 mm, travel 

range 78 x 53 mm 

  Coaxial coarse and fine focusing mechanism with gradua-

tions on the fine focusing mechanism 

  Lighting:  3 W LED (colour  temperature  5000  K),  adjusta-

ble 

  Integrated power supply unit 100...240 V, 50/60 Hz, 3W 

  Fuse: F3.15A/250V 

  Includes dust cover 

  Dimensions (H x W x L): 39 cm x 23 cm x 28 cm 

  Weight: 8.5 kg (MIC-222); 8.6 kg (MIC-231) 

8  WASTE DISPOSAL 

The  packaging  consists  predominately  of  environmentally 
compatible  materials  because  they  can  be  recycled  and 
should  be  passed  on  for  disposal  by  the  local  recycling  ser-
vice. 
 

Should you no longer require this product, 
do not dispose of it with the household re-
fuse.  