3
TECHNICAL DATA
General
Inputs
BNC sockets
Overload protection
±250V in all operating modes
Output
Pair of 4mm safety sockets
Voltage
0...10V
Output resistance
1k
Ω
Overload protection
short-circuit proof
Current measurement
Measurement ranges
100µA. 10µA...10pA (8 ranges)
Accuracy
typ.
≤
3%
Zero-point drift
typ.
≤
0.5%/K
Voltage drop
typ.
≤
0.5mV
Minimum external resistance
Range:
100µA
100
Ω
10µA
1k
Ω
1µA
10k
Ω
100nA
100k
Ω
10nA
1M
Ω
1nA
10M
Ω
100pA
100M
Ω
10pA
1G
Ω
Voltage measurement
Measurement ranges
10V, 1V...100µV (6 ranges)
Accuracy
typ.
≤
3% (ranges 10V...1mV)
typ.
≤
5% (range 100µV)
Zero-point drift
typ.
≤
1%/K (ranges 10V...1mV)
typ.
≤
2%/K (range 100µV)
Internal resistance
>10
11
Ω
Charge measurement
Measurement ranges
1000nAs, 100nAs...0.1nA
(5 ranges)
Accuracy
typ.
≤
3%
Measurement capacitor 100nF ± 1%
Insulation resistance
> 5 · 10
12
Ω
Self-charge rate
typ. < ± 12pAs / min
4
GUARANTEE
We grant a guarantee of 6 months on equipment that we
have supplied; it does not include natural wear and tear and
faults which are the result of improper handling.
The manufacturer can only be regarded as being responsi-
ble for the proper function and safety characteristics of the
equipment if maintenance, repair and modifications have
been carried out by the manufacturer or by agents expres-
sly authorised by the manufacturer.
3
13620.93