1

PURPOSE AND DESCRIPTION

The DC Amplifier is a versatile instrument for the measure-

ment of very low DC currents, for the measurement of char-

ges - in particular in experiments on electrostatics - and also

for the quasi-static measurement of direct voltages.

The selection of the 12 measurement ranges is made using

buttons. A matrix of light emitting diodes indicates the active

measurement range. An output voltage of 10V corresponds

to a full scale deflection. Any voltmeter, recorder or compu-

ter measurement system (COMEX, COBRA, CENT) can be

used for the display.

2

OPERATION

2.1 Initial operation

The DC Amplifier is connected to the AC mains using the

equipment connecting lead supplied with the instrument.

Please ensure that the mains voltage agrees with the ope-

rating voltage details shown on the nameplate. The rectan-

gular fuse holder situated above the equipment connection

plug can only be opened with the aid of a screwdriver or si-

milar tool when the mains lead is withdrawn.

Replacement fuse: M 0.2 C.

The mains switch is located at the back of the unit. The in-

strument is ready for operation immediately after being swit-

ched on.

In order that the measurement accuracy can be maintained,

a warm-up time of about 5 minutes should be allowed be-

fore taking measurements.

2.2 FUNCTIONAL ELEMENTS AND CONTROLS

1.1 BNC socket INPUT I, Q

for connecting the measurement circuit in the opera-

ting modes for current and charge measurement. The

outer pole is connected with the equipment ground

and the protective earth.

1.2 BNC socket INPUT V

for connecting the measurement circuit in the opera-

ting mode for voltage measurement. The outer pole is

connected with the equipment ground and the protec-

tive earth.

2 Button 

I

for switching into the operating mode “Current measu-

rement”.

3 Button 

Q

for switching into the operating mode “Charge measu-

rement”.

4 Button 

V

for switching into the operating mode “Voltage measu-

rement”.

5 Socket 

“Ground”

connected internally with the equipment ground and

the protective earth conductor.

6  Pair of sockets OUTPUT

for the connection of an instrument with a range

10Vdc range.

The internal voltage limit becomes active at an output

voltage of about 10.5V - i.e. the output voltage then

does not correspond to the measurement.

7 Button 

INVERT

for inverting the output signal.

8

Button

for reducing the measurement range in all operating

modes.

PHYWE SYSTEME GMBH · Robert-Bosch-Breite 10 · D-37079 · Göttingen · Telefon (05 51) 6 04-0 ·Telefax (05 51) 60 41 07

13620.93

Operating Instructions

The unit complies 

with the corresponding 

EC guidelines.