background image

4

HANDLING

4.1 Starting up

A smoothed, floating DC supply voltage between 14 and

18V is to be applied to the socket pair 

5

, with the correct

polarity. The voltage source must be capable of being

loaded up to 200 mA; universal power supply 13500.93 or

double power supply 13520.93 are suitable. The necessary

earth contact of the EFM is conveniently connected to the

4 mm transverser hole in the housing stem. This hole can

also be used to tighten the screw fastening of the stem with

the aid of a spanner or the like.

Any continuous voltmeter of adequate measurement range

is to be connected to output 3. Output voltage increases

linearly with field intensity. Depending on the direction of the

field, it will reach +10 V or -10 V for the selected final value

of the measurement range. As an optimal solution for

demonstration purposes, we recommend Analogue-Demo

Multimeter ADM 2, 13820.00 with ±10 V- measurement

range and zero in the centre of the scale. The sign of the

pointer deflection indicates the direction of the field at the

measuring electrode: provided the moving coil instrument is

connected to output 

2

with the correct polarity, a positive

deflection denotes a field towards the measuring electrode,

or a positive potential with respect to equipment earth of the

capacitor plate of the voltage adapter.

Attention:

when the unit is switched on, no intense field

must act on the measuring head, because this would

disturb the automatic function test. Please cover the

measuring head with the protective cap 

8.1

before

switching on the unit, or, in case the voltage measuring

head 

10

is used, short input 

11

. After switching on, the unit

is ready for work as soon as the luminous diodes for the

range display stop flashing. If flashing does not stop after a

few seconds, this shows the function test was disturbed by

electric fields. In this case, briefly switch off power supply

and switch it on again once the measuring head is free of

electric fields.

4.2 Electrical zero adjustment

Before making the measurement, electrical zero adjustment

has still to be carried out, because even if no electric field is

acting on the measuring electrode of the instrument, the

pointer of the moving-coil instrument can deviate from the

set mechanical zero-point (scale centre). The appropriate

correction of the pointer position can be carried out with the

aid of the rotary knob 

2

of the EFM, and most accurately

when the most sensitive measuring range is set by the

push-button 

6

. Naturally, the zero adjustment must be

carried out in the absence of the field to be measured.

It should be noted that, in addition to the measurement field,

stray electrostatic fields – caused by charged parts in the

vicinity – will sometimes affect the measuring head. It is true

that these fields can easily be screened off by means of the

protective cap 

8.1

for adjustment purposes, but during the

actual measurement (without the cap), they would again be

effective. The existence of interference fields can be

recognised from the fact that the pointer deflections differ

considerably with and without the protective cap. As far as

possible, stray electrostatic fields should be avoided. A

slight residual field can be eliminated from the measured

result by the electrical zero adjustment.

A further factor influencing the measurement is the voltaic

potential between the measuring electrode and measuring

apparatus (environment). Of course, its field only becomes

significant when working in the most sensitive measuring

range. It can be eliminated if the electrical zero adjustment

is carried out in the experimental jig itself, which must of

course be in a discharged condition. Practical examples:

For electrostatic voltage measurements, the zero

adjustment is carried out with the voltage adapter 

10

attached and short-circuited with a connection lead. In the

case of field strength measurements in a plate capacitor

similar to the calibration apparatus, with the measuring

head inserted in the hole in one plate (Fig. 2), the zero

adjustment is carried out with the capacitor plates short-

circuited.

With relatively open measuring arrangements, where the

measuring head is not screened from stray fields from the

surroundings, the pointer deflection will also be influenced

by the operator himself – e.g. when operating the rotary

knob 

2

–; therefore one should move sufficiently far away

from the EFM in order to read the pointer position. This

applies both to the zero adjustment and the actual

measurement.

The zero drift of the EFM is relatively small and can be

practically disregarded over the time of performance of an

experiment. However, when working with an open

experimental arrangement, it should be expected that stray

electrostatic fields will vary in an uncontrolled manner or

even come into being in the course of the experiment. It is

therefore advisable to repeat the zero check from time to

time.

4.3 Treatment of the measuring system

The gold-plated measuring system (and also the capacitor

plate of the voltage adapter) should not be contaminated

and also should not be touched by hand. Above all, the

rotating chopper wheel must not be touched. For an

inexperienced observer, the rotation is not directly visible or

audible. The Teflon insultation of the voltage adapter should

also not be touched or contaminated, so that no conductive

surfaces are formed. When not in use, the sensitive

measuring system is to be covered with the protective cap

unless the voltage adapter is fitted.

4.4 Hints for working with the EFM

The EFM will be set up according to the particular

experimental requirements. The surface of the measuring

electrode should be arranged perpendicular to the direction

of the field to be investigated.

Unsatisfactory results in electrostatic experiments are

mainly due, apart from inadequate insulation of stems and

supports, to stray electrostatic fields They can be caused

not only by charged bodies which do not belong to the

actual measurement arrangement (e.g. high-voltage

sources), but also by uncontrollable charges on insulating

plastic surfaces. A few hints on the elimination of these error

sources will be given below.

The high-voltage equipment is to be set up as far away as

possible from the EFM, in order that its field disturbs the

measurement as little as possible. As far as the experiment

permits, the voltage will be switched off during the

measurement; it is generally advisable not to alter the

voltage setting for this.

Charges on the experimental bench can be eliminated or at

least reduced by spraying with an anti-static spray. Live

connectors should be prevented from coming into contact

with the bench surface. On the other hand, when the 50kV

connecting cables, 0.5 m or 1 m long, 07366.00 and

07367.00, are used, their routing is substantially uncritical.

Charges on articles of clothing containing synthetic fibres

are sometimes a very considerable source of stray fields.

Their field becomes noticeable when the wearer comes

near the EFM. It is therefore advisable not to wear such

clothing during electrostatic experiments. In any case, the

operator should remain at an appropriate distance from the

EFM when reading the measured value.

Particular attention is to be paid to the Plexiglass stems on

3

11500.10

Reviews: