background image

 

 

 

 

 

 

www.phywe.com, © All rights reserved 

 

09057-51 / 0414

 

2  PURPOSE AND CHARACTERISTICS 

Each plug-in consists of a pre-adjusted X-ray tube in a 
sheet steel housing with a handle. All plug-ins are sup-
plied ready for use in the basic XR 4.0 X-ray expert 
unit. 

 

Pre-adjusted tubes with quick-change technol-
ogy 

 

Hot parts contact protection  

 

Housing with ratchet and pawl mechanism and 
2 safety contact pins which only unlock for tube 
operation when a plug-in has been correctly 
inserted. 

3  FUNCTIONAL AND OPERATING ELEMENTS 

The following ready for use plug-ins are available: 

 

Plug-in with Cu X-ray tube, article no. 09057-51 

 

Plug-in with Mo X-ray tube, article no. 09057-61 

 

Plug-in with Fe X-ray tube, article no. 09057-71 

 

Plug-in with W X-ray tube, article no. 09057-81 

(For wavelengths and energies of the characteristic X-ray 
lines refer to the Technical specifications and the Appendix). 
Each plug-in consists of a factory-adjusted X-ray tube in a 
sheet steel housing. To prevent damage from overheating, 
each tube is enclosed in a Duran glass cylinder which has 
side tubes for forced air cooling via the fan in the basic de-
vice. 
Each plug-in is equipped with an HV plug and a plug for tube 
cathode heating. Tube operating parameters can so be 
picked up via the corresponding sockets of the basic device.  
Two metal pins at the sides release a safety micro-switch 
which safeguards the tube plug-in chamber on correct inser-
tion of an plug-in. 

4  NOTES ON OPERATION 

Handle with care! 
First start up: 

 

X-ray tubes are not to be operated under full power on their 
first start up. For the running-in of a tube we recommend that 
you first operate it for a period of 10 minutes with a maximum 
beam current but with an acceleration voltage which is not 
above 25 kV. 
This procedure must be repeated when a tube has not been 
used for a period of some weeks. 

 

Minimum life time of the tube: 500 operating hours 

5  HANDLING 

Handle with care! 

Inserting and changing X-ray tube inserts: 

Stop X-ray tube operation before carrying out a change 
of plug-ins. 

Plug-ins  must  be  removed  without  tilting  them.  To  insert  an 
plug-in, fit the guide tabs in the guide rails without tilting the 
plug-in. Ensure that all plug connections have secure contact 
by pushing the plug-in in to the stop. 

6  TECHNICAL SPECIFICATIONS 

 

Target angle                  19° 

 

Max. operating values  1 mA/35 kV-DC 

 

Test voltage          50 kV 

 

Dimensions           (26.7x14.8x20.3) cm 

 

Weight                   4.3 kg 

 

Minimum life time of the tube: 500 operating hours 

 

Characteristic X-ray lines for copper: 

o

 

K-alpha: 8.03 keV; (154,2 pm) 

o

 

K -beta: 8.90 keV; (139,2 pm) 

 

Characteristic X-ray lines for molybdenum: 

o

 

K -alpha: 17.4 keV 

o

 

K -beta: 19.6 keV 

 

Characteristic X-ray lines for iron: 

o

 

K -alpha: 6.40 keV; (194 pm) 

o

 

K -beta: 7.06 keV; (176 pm) 

 

Characteristic X-ray lines for tungsten: 

o

 

See Appendix 

7  PARTS SUPPLIED 

  Complete with dust-protective hood 

8  ACCESSORIES 

09057-01 XR 4.0 X-ray diaphragm tube, d = 1 mm 
09057-02 XR 4.0 X-ray diaphragm tube, d = 2 mm 
09057-03 XR 4.0 X-ray diaphragm tube, d = 5 mm 

 

Fig. 2: Side view of a tube plug-in 

(1) 

Reviews: